ONLINE MEASUREMENT OF SARCOMERE LENGTH FROM DIFFRACTION PATTERNS IN MUSCLE

被引:39
作者
IWAZUMI, T
POLLACK, GH
机构
[1] Department of Anesthesiology, University of Washington, Seattle
关键词
D O I
10.1109/TBME.1979.326514
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Two different systems designed to extract sarcomere length and several other variables from the light diffraction pattern in muscle are described. One system utilizes a linear photodiode array to detect the intensity distribution of diffracted light from muscle. This system provides computed outputs of sarcomere length, first order width, and time rate of sarcomere length change according to two different algorithms. Another system utilizes a Schottky photo-position sensor and provides sarcomere length and integrated first order intensity. The accuracy (within 1%) and resolution (2 nm) of both systems are comparable. The former system offers better convenience of operation while the latter offers wider bandwidth (2.5 kHz vs. 100 kHz). © 1979 IEEE
引用
收藏
页码:86 / 93
页数:8
相关论文
共 8 条
[1]  
BOREJDO J, 1976, Journal of Mechanochemistry and Cell Motility, V3, P155
[2]  
CLEWORTH DR, 1972, J PHYSIOL-LONDON, V227, P1
[3]   OPTICAL DIFFRACTION STUDY OF MUSCLE-FIBERS [J].
FUJIME, S .
BIOCHIMICA ET BIOPHYSICA ACTA, 1975, 379 (01) :227-238
[4]  
HALPERN W, 1977, P SAN DIEGO BIOMED S, V16, P429
[5]   STRUCTURAL CHANGES IN MUSCLE DURING CONTRACTION - INTERFERENCE MICROSCOPY OF LIVING MUSCLE FIBRES [J].
HUXLEY, AF ;
NIEDERGERKE, R .
NATURE, 1954, 173 (4412) :971-973
[6]   CHANGES IN THE CROSS-STRIATIONS OF MUSCLE DURING CONTRACTION AND STRETCH AND THEIR STRUCTURAL INTERPRETATION [J].
HUXLEY, H ;
HANSON, J .
NATURE, 1954, 173 (4412) :973-976
[7]   OPTICAL DIFFRACTION STUDIES OF MUSCLE FIBERS [J].
KAWAI, M ;
KUNTZ, ID .
BIOPHYSICAL JOURNAL, 1973, 13 (09) :857-876
[8]   DUAL-CHANNEL DIFFRACTOMETER UTILIZING LINEAR IMAGE SENSOR CHARGE-COUPLED-DEVICES [J].
PAOLINI, PJ ;
BASKIN, RJ ;
ROOS, KP ;
CLINE, JW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1976, 47 (06) :698-702