Numerical Analysis on the Effect of Wall Shear Stress Around the Ring Drug-Eluting Stent

被引:2
作者
Seo, Taewon [1 ]
Barakat, Abdul [1 ]
机构
[1] Univ Calif Davis, Dept Mech & Aeronaut Engn, Davis, CA 95616 USA
关键词
Drug-eluting Stent; Atherosclerosis; Drug Concentration; Wall Shear Stress;
D O I
10.3795/KSME-B.2007.31.1.021
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The use of drug-eluting stents has dramatically reduced the incidence of restenosis however, much remains to be learned about the performance of these stents. In the present study, we tested the hypothesis that the design of drug-eluting stents influences the efficacy of local drug delivery to the arterial wall and that this effect depends on both arterial geometry and the prevailing flow conditions. We performed computational simulations in which the coupled Navier-Stokes and advection-diffusion equations were solved to determine the flow field and drug concentration in the vicinity of model drug-eluting stents It is found that the characteristics of flow phenomena can be influenced greatly by the ratio of stent diameter to vessel diameter. The presence of drug-eluting stent may have profound effect on wall shear stresses, recirculation sizes and drug distributions. The results show that recirculation zone is influenced by the imposed flow conditions and stent diameter. In pulsatile flow, the low wall shear stress and high drug concentration occur along the arterial wall during the decelerating flow conditions. These results could provide the guideline for future drug-eluting stent designs toward reducing restenosis by affecting local wall shear stress distributions associated with neointimal hyperplasia.
引用
收藏
页码:21 / 28
页数:8
相关论文
共 13 条
  • [1] Experimental and computational flow evaluation of coronary stents
    Berry, JL
    Santamarina, A
    Moore, JE
    Roychowdhury, S
    Routh, WD
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2000, 28 (04) : 386 - 398
  • [2] Cohen David J., PERSPECTIVES EC DRUG
  • [3] Stent and artery geometry determine intimal thickening independent of arterial injury
    Garasic, JM
    Edelman, ER
    Squire, JC
    Seifert, P
    Williams, MS
    Rogers, C
    [J]. CIRCULATION, 2000, 101 (07) : 812 - 818
  • [4] Pulsatile flow in the human left coronary artery bifurcation: Average conditions
    He, XJ
    Ku, DN
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (01): : 74 - 82
  • [5] Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study
    Hoffmann, R
    Mintz, GS
    Dussaillant, GR
    Popma, JJ
    Pichard, AD
    Satler, LF
    Kent, KM
    Griffin, J
    Leon, MB
    [J]. CIRCULATION, 1996, 94 (06) : 1247 - 1254
  • [6] TIME-COURSE OF RESTENOSIS DURING THE 1ST YEAR AFTER EMERGENCY CORONARY STENTING
    KASTRATI, A
    SCHOMIG, A
    DIETZ, R
    NEUMANN, FJ
    RICHARDT, G
    [J]. CIRCULATION, 1993, 87 (05) : 1498 - 1505
  • [7] Ku D. N., 1994, ATHEROSCLEROSIS, V5, P293
  • [8] RESTENOSIS AFTER CORONARY ANGIOPLASTY - POTENTIAL BIOLOGIC DETERMINANTS AND ROLE OF INTIMAL HYPERPLASIA
    LIU, MW
    ROUBIN, GS
    KING, SB
    [J]. CIRCULATION, 1989, 79 (06) : 1374 - 1387
  • [9] Computational study of fluid mechanical disturbance induced by endovascular stents
    Seo, T
    Schachter, LG
    Barakat, AI
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2005, 33 (04) : 444 - 456
  • [10] Stent Design Using Computational Fluid Dynamics
    Seo, Taewon
    Kim, Taedong
    Barakat, Abdul
    [J]. TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2005, 29 (09) : 1042 - 1048