EXPERIMENTAL-VERIFICATION OF A 3-DIMENSIONAL DOSE CALCULATION ALGORITHM USING A SPECIALLY DESIGNED HETEROGENEOUS PHANTOM

被引:5
作者
AYYANGAR, K [1 ]
PALTA, JR [1 ]
SWEET, JW [1 ]
SUNTHARALINGAM, N [1 ]
机构
[1] THOMAS JEFFERSON UNIV,DEPT RADIAT ONCOL & NUCL MED,PHILADELPHIA,PA 19107
关键词
3-D TREATMENT PLANNING; DOSE CALCULATION; PHOTON BEAM DOSIMETRY; INHOMOGENEITY CORRECTIONS;
D O I
10.1118/1.597072
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A solid heterogeneous phantom made up of 25- and 50-mm cubes of materials with different electron densities was used to verify the accuracy of a three-dimensional (3-D) dose calculation algorithm. This algorithm uses 3-D information obtained from contiguous CT (computed tomography) slices, spaced 5 mm apart. Primary and scatter doses at a point are calculated by using information from ray-tracing CT voxels. The algorithm was developed on a Stardent model 1500 Supergraphic workstation. Cubes of materials with different electron densities were stacked up to simulate finite heterogeneities in three dimensions. This design allows verification of the algorithm for surface contour corrections and finite heterogeneities in the treatment field. Thermoluminescent lithium fluoride chips were placed in grooves milled on the cubes for dose measurement at various points. Different experiments were performed to investigate both the accuracy of the dose calculation algorithm and the utility of the versatile test phantom.
引用
收藏
页码:325 / 329
页数:5
相关论文
共 10 条
[1]   OPTIMIZATION OF PARAMETERS FOR FITTING LINEAR-ACCELERATOR PHOTON BEAMS USING A MODIFIED CBEAM MODEL [J].
AYYANGAR, K ;
DAFTARI, I ;
PALTA, J ;
SUNTHARALINGAM, N .
MEDICAL PHYSICS, 1989, 16 (06) :896-901
[2]  
AYYANGAR K, 1990, MED PHYS, V17, P513
[3]  
BATTISTA JJ, 1989, PHYS MED BIOL, V26, P81
[4]   THE VARIATION OF SCATTERED X-RAYS WITH DENSITY IN AN IRRADIATED BODY [J].
OCONNOR, JE .
PHYSICS IN MEDICINE AND BIOLOGY, 1957, 1 (04) :352-369
[5]   IMPLICATIONS OF COMPUTED TOMOGRAPHY FOR INHOMOGENEITY CORRECTIONS IN PHOTON BEAM DOSE CALCULATIONS [J].
SONTAG, MR ;
BATTISTA, JJ ;
BRONSKILL, MJ ;
CUNNINGHAM, JR .
RADIOLOGY, 1977, 124 (01) :143-149
[6]  
SONTAG MR, 1978, RADIOLOGY, V129, P789
[7]   VERIFICATION BY MONTE-CARLO METHODS OF A POWER LAW TISSUE AIR RATIO ALGORITHM FOR INHOMOGENEITY CORRECTIONS IN PHOTON-BEAM DOSE CALCULATIONS [J].
WEBB, S ;
FOX, RA .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (02) :225-240
[8]   ON METHODS OF INHOMOGENEITY CORRECTIONS FOR PHOTON TRANSPORT [J].
WONG, JW ;
PURDY, JA .
MEDICAL PHYSICS, 1990, 17 (05) :807-814
[9]   A NEW APPROACH TO CT PIXEL-BASED PHOTON DOSE CALCULATIONS IN HETEROGENEOUS MEDIA [J].
WONG, JW ;
HENKELMAN, RM .
MEDICAL PHYSICS, 1983, 10 (02) :199-208
[10]   EXTENDING THE CONCEPT OF PRIMARY AND SCATTER SEPARATION TO THE CONDITION OF ELECTRONIC DISEQUILIBRIUM [J].
WOO, MK ;
CUNNINGHAM, JR ;
JEZIORANSKI, JJ .
MEDICAL PHYSICS, 1990, 17 (04) :588-595