STEREOTACTIC HEAD FRAME FOR USE WITH CT BODY SCANNERS

被引:111
作者
BROWN, RA
机构
关键词
CT; Neurosurgical localization; Stereotaxis;
D O I
10.1097/00004424-197907000-00006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A prototype stereotactic head frame has been constructed for use in conjunction with a computed tomographic (CT) body scanner. Interactive software has been developed to allow target delineation using a standard CT scanner computer and image display console. The mathematic equations fundamental to this software are described. In preliminary experiments, it has been demonstrated that this frame can accurately direct a probe to an appropriately delineated target in a series of CT scans. A theoretic method for determining accuracy is presented; results suggest that this system represents a significant improvement over classic methods of neurosurgical stereotactic localization. © J. B. Lippincott Company.
引用
收藏
页码:300 / 304
页数:5
相关论文
共 4 条
  • [1] STEREOTAXIC COMPUTED TOMOGRAPHY
    BERGSTROM, M
    GREITZ, T
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1976, 127 (01) : 167 - 170
  • [3] POTENTIAL ARTIFACTS ASSOCIATED WITH SCANNING PATTERN OF EMI SCANNER
    GOODENOUGH, DJ
    WEAVER, KE
    DAVIS, DO
    [J]. RADIOLOGY, 1975, 117 (03) : 615 - 619
  • [4] SORENSON J A, 1979, Medical Physics (Woodbury), V6, P68, DOI 10.1118/1.594554