CELLULAR INTERACTIONS WITH BIOMATERIALS - INVIVO CRACKING OF PRE-STRESSED PELLETHANE 2363-80A

被引:119
作者
ZHAO, Q
AGGER, MP
FITZPATRICK, M
ANDERSON, JM
HILTNER, A
STOKES, K
URBANSKI, P
机构
[1] CASE WESTERN RESERVE UNIV,DEPT BIOMED ENGN,CLEVELAND,OH 44106
[2] CASE WESTERN RESERVE UNIV,DEPT PATHOL,CLEVELAND,OH 44106
[3] MEDTRONIC INC,MINNEAPOLIS,MN 55432
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1990年 / 24卷 / 05期
关键词
D O I
10.1002/jbm.820240508
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The phenomenon of stress cracking of Pellethane 2363‐80A (PEU) was investigated using the cage implant system. A cytotoxic polyvinylchloride (PVC) and a silicone rubber containing an anti‐inflammatory steroid were used to create inflammatory environments in which the biostability of the pre‐stressed PEU was tested. These coimplants provided alternative in vivo environments to study in vivo polymer interactions. The inflammatory responses to the implanted cages were monitored by analyzing the exudates aspirated from the cages at different implantation times over 21 days. The pre‐stressed PEU specimens were retrieved after 5, 10, and 15 weeks postimplantation and examined by optical microscopy (OM) and scanning electron microscopy (SEM). The results support the conclusion that in vivo cracking of stressed (strained) Pellethane 80A is related to cell–polymer interactions. Severe cracking or rupture of the implanted PEU specimens was observed as early as 5 weeks postimplantation. Molecular chain degradation of the implanted specimens was evident from molecular weight measurements. Neither surface cracking nor degradation of macromolecules was found on the pre‐stressed PEU specimens with the added cytotoxic PVC implanted over 15 weeks. No cracking was observed on the pre‐stressed specimens in the presence of steroid silicone rubber, even after 10 weeks implantation. Copyright © 1990 John Wiley & Sons, Inc.
引用
收藏
页码:621 / 637
页数:17
相关论文
共 18 条
  • [1] ANDERSON JM, 1984, BIOMATERIALS, V5, P5, DOI 10.1016/0142-9612(84)90060-7
  • [2] ANDERSON JM, 1988, ASAIO, V11, P101
  • [3] HENSON PM, 1980, AM J PATHOL, V101, P494
  • [4] LEMM M, 1984, POLYURETHANES BIOMED, P103
  • [5] INVIVO BIOCOMPATIBILITY STUDIES .1. THE CAGE IMPLANT SYSTEM AND A BIODEGRADABLE HYDROGEL
    MARCHANT, R
    HILTNER, A
    HAMLIN, C
    RABINOVITCH, A
    SLOBODKIN, R
    ANDERSON, JM
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (02): : 301 - 325
  • [6] INVIVO BIOCOMPATIBILITY STUDIES .5. INVIVO LEUKOCYTE INTERACTIONS WITH BIOMER
    MARCHANT, RE
    MILLER, KM
    ANDERSON, JM
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (09): : 1169 - 1190
  • [7] INVIVO BIOCOMPATIBILITY STUDIES .2. BIOMER - PRELIMINARY CELL-ADHESION AND SURFACE CHARACTERIZATION STUDIES
    MARCHANT, RE
    ANDERSON, JM
    PHUA, K
    HILTNER, A
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (03): : 309 - 315
  • [8] CELL-ADHESION AND INTERACTIONS WITH BIOMATERIALS
    MARCHANT, RE
    [J]. JOURNAL OF ADHESION, 1986, 20 (03) : 211 - 225
  • [9] INVIVO BIOCOMPATIBILITY STUDIES .7. INFLAMMATORY RESPONSE TO POLYETHYLENE AND TO A CYTOTOXIC POLYVINYLCHLORIDE
    MARCHANT, RE
    ANDERSON, JM
    DILLINGHAM, EO
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1986, 20 (01): : 37 - 50
  • [10] PARINS DJ, 1981, IN VIVO DEGRADATION