ARTERIAL OXYGEN STATUS DETERMINED WITH ROUTINE PH/BLOOD GAS EQUIPMENT AND MULTIWAVELENGTH HEMOXIMETRY - REFERENCE VALUES, PRECISION, AND ACCURACY

被引:35
作者
SIGGAARDANDERSEN, O
WIMBERLEY, PD
FOGHANDERSEN, N
GOTHGEN, IH
机构
[1] UNIV COPENHAGEN, GENTOFFE HOSP, DEPT ANESTHESIOL 8 INTENS CARE, DK-2900 HELLERUP, DENMARK
[2] HERNING HOSP, DEPT CLIN CHEM, DK-7400 HERNING, DENMARK
关键词
2,3-DIPHOSPHOGLYCERATE; HEMOGLOBIN-OXYGEN AFFINITY; OXYGEN SATURATION; OXYGEN TENSION;
D O I
10.3109/00365519009087492
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
We measured pH, pCO2, pO2, oxygen saturation, total hemoglobin concentration, and fractions of carboxy- and methemoglobin in arterial blood samples from 35 healthy adults. We used a new algorithm to calculate active hemoglobin concentration, total oxygen concentration, actual half-saturation tension, 2,3-diphosphoglycerate concentration, estimated functional shunt, oxygen extraction tension p(x) (for extracting 2.3 mmol of oxygen per liter of blood, values below 4.5 kPa indicating risk of tissue hypoxia), and the oxygen compensation factor Q(x) (the factor by which the cardiac output should rise to maintain a normal mixed venous pO2 of 5.0 kPa, factors above 1.5 indicating an extra burden on the heart). Analytical precision was evaluated by duplicate determinations. The accuracy of the half-saturation tension was evaluated by comparison with values for simultaneously drawn venous blood, the accuracy of the calculated concentration of 2,3-diphosphoglycerate by comparison with direct enzymatic measurements. We conclude that all the variables may be determined with sufficient accuracy and precision in healthy adults, provided the oxygen saturation is less than 0.97 and the measurements are performed according to the highest state of the art.
引用
收藏
页码:57 / 66
页数:10
相关论文
共 17 条