DIRECT ANALYSIS OF KRYPTOFIX 2.2.2 IN 18FDG BY GAS-CHROMATOGRAPHY USING A NITROGEN-SELECTIVE DETECTOR

被引:18
|
作者
FERRIERI, RA
SCHLYER, DJ
ALEXOFF, DL
FOWLER, JS
WOLF, AP
机构
[1] Department of Chemistry, Brookhaven National Laboratory, Upton
来源
NUCLEAR MEDICINE AND BIOLOGY | 1993年 / 20卷 / 03期
关键词
D O I
10.1016/0969-8051(93)90061-X
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method is described for the direct measurement of Kryptofix 2.2.2 levels in F-18DG preparations dissolved in isotonic saline using gas chromatography coupled with a nitrogen-selective detector. The cryptand can be eluted intact in 2 min time from a short megabore capillary column, and detected at levels as low as 0.25 mug/mL. This method offers rapid, quantitative and sensitive detection of Kryptofix 2.2.2 levels in F-18DG thus making it amenable to routine quality assurance testing.
引用
收藏
页码:367 / 369
页数:3
相关论文
共 50 条