VENOUS CARBON-DIOXIDE EMBOLISM IN PIGS - AN EVALUATION OF END-TIDAL CARBON-DIOXIDE, TRANSESOPHAGEAL ECHOCARDIOGRAPHY, PULMONARY-ARTERY PRESSURE, AND PRECORDIAL AUSCULTATION AS MONITORING MODALITIES

被引:2
|
作者
COUTURE, P
BOUDREAULT, D
DEROUIN, M
ALLARD, M
LEPAGE, Y
GIRARD, D
BLAISE, G
机构
来源
ANESTHESIA AND ANALGESIA | 1994年 / 79卷 / 05期
关键词
D O I
暂无
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We evaluated the effects of CO2 embolism on end-tidal carbon dioxide (ETCO(2)) and compared four methods for detection of gas embolism. Fourteen pigs were monitored for CO2 embolism with transesophageal echocardiography (TEE), changes in ETCO(2) changes in mean pulmonary artery pressure (MPAP), and precordial auscultation (AUSC). Serial injections of CO2 (ranging from 0.05 to 5.0 mL/kg) were performed in seven pigs (Group 1). In the other seven pigs, CO2 was infused at rates between 0.01 and 0.4 mt kg(-1) min(-1) (Group 2). Positive responses were defined as an acute change in heart sounds (AUSC), visualization of gas bubbles in the right cardiac chambers on TEE, an increase in MPAP greater than or equal to 3 mm Hg, and an acute change (increase or decrease) in ETCO(2) greater than or equal to 3 mm Hg. In both groups, positive responses to CO2 embolism were represented by an initial decrease in ETCO(2) The frequency with which positive responses were observed revealed that TEE was the most sensitive method (P < 0.05), whereas no differences were found among the other methods. In conclusion, in this model, positive response to CO2 embolism was represented by a decrease in ETCO(2). TEE was the most sensitive method of detection of CO2 embolism, and ETCO(2), MPAP, and AUSC were equally sensitive.
引用
收藏
页码:867 / 873
页数:7
相关论文
共 50 条
  • [1] A COMPARISON OF THE SENSITIVITY OF PULMONARY-ARTERY PRESSURE, END-TIDAL CARBON-DIOXIDE, AND END-TIDAL NITROGEN IN THE DETECTION OF VENOUS AIR-EMBOLISM IN THE DOG
    DRUMMOND, JC
    PRUTOW, RJ
    SCHELLER, MS
    ANESTHESIA AND ANALGESIA, 1985, 64 (07): : 688 - 692
  • [2] MONITORING OF END-TIDAL CARBON-DIOXIDE TO DETECT VENOUS AIR-EMBOLISM
    PAUL, WL
    MUNSON, ES
    BRITISH JOURNAL OF ANAESTHESIA, 1981, 53 (03) : P313 - P314
  • [3] NONINVASIVE MONITORING OF CARBON-DIOXIDE - A COMPARISON OF THE PARTIAL-PRESSURE OF TRANS-CUTANEOUS AND END-TIDAL CARBON-DIOXIDE WITH THE PARTIAL-PRESSURE OF ARTERIAL CARBON-DIOXIDE
    PHAN, CQ
    TREMPER, KK
    LEE, SE
    BARKER, SJ
    JOURNAL OF CLINICAL MONITORING, 1987, 3 (03): : 149 - 154
  • [4] END TIDAL CARBON-DIOXIDE MONITORING FOR DETECTION OF VENOUS AIR-EMBOLISM
    MULEY, SS
    SAINI, SS
    DASH, HH
    BITHAL, PK
    INDIAN JOURNAL OF MEDICAL RESEARCH SECTION B-BIOMEDICAL RESEARCH OTHER THAN INFECTIOUS DISEASES, 1990, 92 : 362 - 366
  • [5] END-TIDAL CARBON-DIOXIDE AS AN INDICATOR OF ARTERIAL CARBON-DIOXIDE IN NEUROINTENSIVE CARE PATIENTS
    RUSSELL, GB
    GRAYBEAL, JM
    JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 1992, 4 (04) : 245 - 249
  • [6] ARTERIAL TO END-TIDAL CARBON-DIOXIDE GRADIENTS
    SOSIS, M
    ANESTHESIA AND ANALGESIA, 1988, 67 (05): : 486 - 486
  • [7] The detection of carbon dioxide embolism during laparoscopy in pigs: A comparison of transesophageal Doppler and end-tidal carbon dioxide monitoring
    Mann, C
    Boccara, G
    Fabre, JM
    Grevy, V
    Colson, P
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1997, 41 (02) : 281 - 286
  • [8] COLORIMETRIC END-TIDAL CARBON-DIOXIDE MONITORING FOR TRACHEAL INTUBATION
    GOLDBERG, JS
    RAWLE, PR
    ZEHNDER, JL
    SLADEN, RN
    ANESTHESIA AND ANALGESIA, 1990, 70 (02): : 191 - 194
  • [9] COLORIMETRIC END-TIDAL CARBON-DIOXIDE DETECTOR
    BHENDE, MS
    PEDIATRIC EMERGENCY CARE, 1995, 11 (01) : 58 - 61
  • [10] A SPURIOUSLY LOW END-TIDAL CARBON-DIOXIDE
    HURLEY, MR
    PAULL, JD
    ANAESTHESIA AND INTENSIVE CARE, 1991, 19 (04) : 615 - 616