AVERAGED STEADY-STATE VISUAL EVOKED CORTICAL POTENTIALS AT ARTIFICIALLY RAISED INTRAOCULAR-PRESSURE

被引:3
作者
KREMMER, S
STODTMEISTER, R
TOLKSDORF, A
PILLUNAT, LE
机构
[1] Department of Ophthalmology, University of Ulm
关键词
ARTIFICIALLY INCREASED INTRAOCULAR PRESSURE; AVERAGER; GLAUCOMA; PRESSURE TOLERANCE TEST; STEADY-STATE VISUAL EVOKED CORTICAL POTENTIALS;
D O I
10.1007/BF00156008
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
By recording steady-state visual evoked cortical potentials while intraocular pressure is artificially increased, information can be obtained on the pressure tolerance of the optic nerve head. Such experiments have previously been performed by a vector voltmeter technique. We studied the visual evoked cortical potentials in 30 healthy volunteers with artificially increased intraocular pressure, but we used an averager instead of a vector voltmeter. The results were similar except that the noise level in averaging was higher than with the vector voltmeter technique. This observation confirms that the signal-to-noise ratio is much better with the vector voltmeter technique than with the averaging technique. Our results show that averaging can be used in pressure tolerance testing, but the amplitude cannot be observed as far down as in the vector voltmeter technique. This limits the clinical value of averagers in this application.
引用
收藏
页码:189 / 196
页数:8
相关论文
共 15 条
[1]  
BATSCHELET F, 1971, INTRO MATH LIFE SCI, P124
[2]  
Hayreh SS, 1978, GLAUCOMA CONCEPTIONS, P78
[3]  
KAUFMANN M, 1990, Fortschritte der Ophthalmologie, V87, P32
[4]  
Klem G H, 1999, Electroencephalogr Clin Neurophysiol Suppl, V52, P3
[5]   HIRNPOTENTIALANDERUNGEN BEI WILLKURBEWEGUNGEN UND PASSIVEN BEWEGUNGEN DES MENSCHEN - BEREITSCHAFTSPOTENTIAL UND REAFFERENTE POTENTIALE [J].
KORNHUBE.HH ;
DEECKE, L .
PFLUGERS ARCHIV FUR DIE GESAMTE PHYSIOLOGIE DES MENSCHEN UND DER TIERE, 1965, 284 (01) :1-&
[6]  
KUKAN F, 1931, KLIN MONATSBL AUGENH, V89, P553
[7]   AUTO-REGULATION OF OCULAR BLOOD-FLOW DURING CHANGES IN INTRAOCULAR-PRESSURE - PRELIMINARY-RESULTS [J].
PILLUNAT, LE ;
STODTMEISTER, R ;
WILMANNS, I ;
CHRIST, T .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1985, 223 (04) :219-223
[8]   PRESSURE COMPLIANCE OF THE OPTIC-NERVE HEAD IN LOW TENSION GLAUCOMA [J].
PILLUNAT, LE ;
STODTMEISTER, R ;
WILMANNS, I .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1987, 71 (03) :181-187
[9]  
Regan D., 1989, HUMAN BRAIN ELECTROP
[10]  
REGAN D, 1982, ANN NY ACAD SCI, V338, P45