AP versus PA positioning in lumbar spine computed radiography: Image quality and individual organ doses

被引:22
作者
Davey, Enda [1 ]
England, Andrew [1 ]
机构
[1] Univ Salford, Directorate Radiog, Manchester M5 4WT, Lancs, England
关键词
Image quality; Positioning; Dose optimisation; Lumbar spine; Computed radiography;
D O I
10.1016/j.radi.2014.11.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Radiological imaging examinations must be optimised in order to ensure that the radiation dose is kept as low as reasonably possible (ALARP). The aim of this study was to compare anteroposterior (AP) and posteroanterior (PA) projections of the lumbar spine, at various kVp increments, in order to establish optimum parameters. Methods: An anthropomorphic phantom was imaged in both the AP/PA projections and at various kVp increments. Acquisitions were undertaken using a Wolverson Acroma X-ray unit and processed using an Agfa computed radiography (CR) unit. The entrance surface dose was recorded and converted to effective and organ doses using PCXMC 2.0 software. Five observers were then asked to evaluate the images, using a two-alternative force choice (2AFC) approach and a scale based on EC guidelines. Results: The PA projection lowered the mean effective dose by 19.8% and also the mean absorbed dose to the stomach (70.4%), colon (61.1%), remainder tissues (33.2%), ovaries (7.3%) and testes (15.9%). However, this was at the expense of slightly inferior image quality, not statistically significant. For AP projections, a higher kVp is a further option for dose reduction. Conclusion: Dose optimisation requires the production of an image that is acceptable for the purpose intended. Based on ALARP, and when taking into consideration the dose reductions in this study, it may now be time to routinely use PA projections when imaging the lumbar spine. The use of a higher kVp should also be considered as an option but would be more useful for AP projections. (C) 2014 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:188 / 196
页数:9
相关论文
共 29 条
  • [1] (AAPM) AAoPiM, 2010, TASK GROUP 18 ASS DI
  • [2] The use of reference image criteria in X-ray diagnostics:: an application for the optimisation of lumbar spine radiographs
    Almén, A
    Tingberg, A
    Besjakov, J
    Mattsson, S
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (09) : 1561 - 1567
  • [3] Lumbar spine radiology: analysis of the posteroanterior projection
    Brennan, PC
    Madigan, E
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (07) : 1197 - 1201
  • [4] Brindhaban A., 2005, Australasian Physical & Engineering Sciences in Medicine, V28, P216
  • [5] Reduction of radiation risks in patients undergoing some X-ray examinations by using optimal projections: A Monte Carlo program-based mathematical calculation
    Chaparian, A.
    Kanani, A.
    Baghbanian, M.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2014, 39 (01) : 32 - 39
  • [6] The effect of patient shield position on gonad dose during lumbar spine radiography
    Clancy, Conor L.
    O'Reilly, Geraldine
    Brennan, Patrick C.
    McEntee, Mark F.
    [J]. RADIOGRAPHY, 2010, 16 (02) : 131 - 135
  • [7] CoE Communities, 1996, EUR GUID QUAL CRIT D
  • [8] Do CEC guidelines under-utilise the full potential of increasing kVp as a dose-reducing tool?
    Doherty, P
    O'Leary, D
    Brennan, PC
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (08) : 1992 - 1999
  • [9] Fauber T.L., 2013, RADIOGRAPHIC IMAGING, V4th ed.
  • [10] Fauber TL, 2004, RADIOGRAPHIC IMAGING