AN EXPERIMENTAL-STUDY IN ELECTRICAL-IMPEDANCE TOMOGRAPHY USING BACKPROJECTION RECONSTRUCTION

被引:34
|
作者
GUARDO, R
BOULAY, C
MURRAY, B
BERTRAND, M
机构
[1] Institut de Génie Biomedical, Ecole Polytechnique et Université de Montréal, Montréal, Québec
关键词
APPLIED POTENTIAL TOMOGRAPHY;
D O I
10.1109/10.83560
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports on experiments designed to evaluate the performance of the equipotentials backprojection method under conditions modeling those of proposed applications of electrical impedance tomography. Small spherical targets were placed inside a saline-filled tank with dimensions similar to a human torso. Data were acquired with a computer-based instrument that applies current to pairs of electrodes located on two horizontal planes and records potential differences between electrodes of a third plane. The relative contrast produced by nonconducting spheres in a uniform saline background was measured on the reconstructed images and used to determine system sensitivity to target volume and to the radial and vertical positions of single spheres. Results show that for radial positions within a critical radius sensitivity is always maximum when the spheres center is on the recording plane and decreases gradually when the target is moved outside this plane. Localization of simple targets in 3-D, with data acquired from multiple recording planes, appears feasible. The results provide guidelines for the interpretation of images with complex 3-D conductivity distributions.
引用
收藏
页码:617 / 627
页数:11
相关论文
共 50 条
  • [21] ELECTRICAL-IMPEDANCE TOMOGRAPHY - THE CONSTRUCTION AND APPLICATION TO PHYSIOLOGICAL MEASUREMENT OF ELECTRICAL-IMPEDANCE IMAGES
    BROWN, BH
    BARBER, DC
    MEDICAL PROGRESS THROUGH TECHNOLOGY, 1987, 13 (02) : 69 - 75
  • [22] IMAGING THE COMPLEX IMPEDANCE IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    JOSSINET, J
    TRILLAUD, C
    CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 : 47 - 50
  • [23] EXPERIMENTAL-STUDY ON ELECTRICAL-IMPEDANCE OF BONE AND EFFECT OF DIRECT-CURRENT ON HEALING OF FRACTURES
    STEFAN, S
    SANSEN, W
    MULIER, JC
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1976, (120) : 264 - 267
  • [24] ELECTRICAL-IMPEDANCE TOMOGRAPHY USING INDUCED AND INJECTED CURRENTS
    GENCER, NG
    IDER, YZ
    KUZUOGLU, M
    CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 : 95 - 99
  • [25] MEASURING LUNG RESISTIVITY USING ELECTRICAL-IMPEDANCE TOMOGRAPHY
    WOO, EJ
    HUA, P
    WEBSTER, JG
    TOMPKINS, WJ
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (07) : 756 - 760
  • [26] EXPERIMENTAL EVALUATION OF IMAGE PLANE SELECTIVITY IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    GUARDO, R
    BOULAY, C
    MURRAY, B
    BERTRAND, M
    IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 : 468 - 469
  • [27] IDENTIFYING ESOPHAGEAL CONTENTS USING ELECTRICAL-IMPEDANCE TOMOGRAPHY
    EROL, RA
    MANGNALL, Y
    LEATHARD, AD
    SMALLWOOD, RH
    BROWN, BH
    CHERIAN, P
    BARDHAN, KD
    PHYSIOLOGICAL MEASUREMENT, 1995, 16 (04) : 253 - 261
  • [28] A CIRCULAR FINITE-ELEMENT MODEL RECONSTRUCTION IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    MOESNER, LN
    ANDERSEN, OT
    DAWIDS, S
    SORENSEN, JA
    IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 : 446 - 447
  • [29] A DIRECT SENSITIVITY MATRIX APPROACH FOR FAST RECONSTRUCTION IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    MORUCCI, JP
    MARSILI, PM
    GRANIE, M
    SHI, Y
    LEI, M
    DAI, WW
    PHYSIOLOGICAL MEASUREMENT, 1994, 15 : A107 - A114
  • [30] A RECONSTRUCTION ALGORITHM OF ELECTRICAL-IMPEDANCE TOMOGRAPHY WITH OPTIMAL CONFIGURATION OF THE DRIVEN ELECTRODES
    KYRIACOU, GA
    KOUKOURLIS, CS
    SAHALOS, JN
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (03) : 430 - 438