EFFECT OF PROSTHETIC AORTIC-VALVE DESIGN ON THE DOPPLER-CATHETER GRADIENT CORRELATION - AN INVITRO STUDY OF NORMAL ST-JUDE, MEDTRONIC-HALL, STARR-EDWARDS AND HANCOCK VALVES

被引:103
作者
BAUMGARTNER, H
KHAN, S
DEROBERTIS, M
CZER, L
MAURER, G
机构
[1] UNIV CALIF LOS ANGELES,SCH MED,LOS ANGELES,CA 90024
[2] CEDARS SINAI MED CTR,DIV CARDIOL & CARDIOVASC SURG,LOS ANGELES,CA 90048
关键词
D O I
10.1016/0735-1097(92)90486-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To evaluate the normal range of Doppler-derived velocities and gradients, their relation to direct flow measurements and the importance of prosthetic valve design on the relation between Doppler and catheter-derived gradients, five sizes of normal St. Jude bileaflet, Medtronic-Hall tilting disc, Starr-Edwards caged ball and Hancock bioprosthetic aortic valves were studied with use of a pulsatile flow model. A strong linear correlation between peak velocity and peak flow, and mean velocity and mean flow, was found in all four valve types (r = 0.96 to 0.99). In small St. Jude and Hancock valves, Doppler velocities and corresponding gradients increased dramatically with increasing flow, resulting in velocities and gradients as high as 4.7 m/s and 89 mm Hg, respectively. The ratio of velocity across the valve to velocity in front of the valve (velocity ratio) was independent of flow in all St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves when the two lowest flow rates were excluded for Hancock valves. Although Doppler peak and mean gradients correlated well with catheter peak and mean gradients in all four valve types, the actual agreement between the two techniques was acceptable only in Hancock and Medtronic-Hall valves. For St. Jude and Starr-Edwards valves, Doppler gradients significantly and consistently exceeded catheter gradients with differences as great as 44 mm Hg. Thus, Doppler velocities and gradients across normal prothetic heart valves are highly flow dependent. However, the velocity ratio is independent of flow. Doppler and catheter gradients across prostheses show acceptable agreement in Medtronic-Hall and Hancock valves, but Doppler gradients may significantly exceed catheter gradients in St. Jude valves and Starr-Edwards valves, especially at high flow rates. The discrepancy between Doppler and catheter gradients in both St. Jude and Starr-Edwards valves is of potential clinical importance and suggests that localized high velocity jets and pressure recovery occur in these valves.
引用
收藏
页码:324 / 332
页数:9
相关论文
共 31 条
  • [1] ARABIA F A, 1989, Biomedical Instrumentation and Technology, V23, P205
  • [2] DISCREPANCIES BETWEEN DOPPLER AND CATHETER GRADIENTS IN AORTIC PROSTHETIC VALVES INVITRO - A MANIFESTATION OF LOCALIZED GRADIENTS AND PRESSURE RECOVERY
    BAUMGARTNER, H
    KHAN, S
    DEROBERTIS, M
    CZER, L
    MAURER, G
    [J]. CIRCULATION, 1990, 82 (04) : 1467 - 1475
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] CONTINUOUS WAVE DOPPLER ECHOCARDIOGRAPHIC MEASUREMENT OF PROSTHETIC VALVE GRADIENTS - A SIMULTANEOUS DOPPLER-CATHETER CORRELATIVE STUDY
    BURSTOW, DJ
    NISHIMURA, RA
    BAILEY, KR
    REEDER, GS
    HOLMES, DR
    SEWARD, JB
    TAJIK, AJ
    [J]. CIRCULATION, 1989, 80 (03) : 504 - 514
  • [5] HYDRAULIC ESTIMATION OF STENOTIC ORIFICE AREA - A CORRECTION OF THE GORLIN FORMULA
    CANNON, SR
    RICHARDS, KL
    CRAWFORD, M
    [J]. CIRCULATION, 1985, 71 (06) : 1170 - 1178
  • [6] DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION
    CHAFIZADEH, ER
    ZOGHBI, WA
    [J]. CIRCULATION, 1991, 83 (01) : 213 - 223
  • [7] EVALUATION OF NORMAL PROSTHETIC VALVE FUNCTION BY DOPPLER ECHOCARDIOGRAPHY
    COOPER, DM
    STEWART, WJ
    SCHIAVONE, WA
    LOMBARDO, HP
    LYTLE, BW
    LOOP, FD
    SALCEDO, EE
    [J]. AMERICAN HEART JOURNAL, 1987, 114 (03) : 576 - 582
  • [8] CURRIE P, 1985, CIRCULATION, P1162
  • [9] DOPPLER ULTRASOUND OF NORMALLY FUNCTIONING MECHANICAL MITRAL AND AORTIC-VALVE PROSTHESES
    GIBBS, JL
    WHARTON, GA
    WILLIAMS, GJ
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1988, 18 (03) : 391 - 398
  • [10] HEGRENAES L, 1985, BRIT HEART J, V54, P396