DETERMINATION OF TERFENADINE AND TERFENADINE ACID METABOLITE IN PLASMA USING SOLID-PHASE EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION

被引:67
作者
COUTANT, JE
WESTMARK, PA
NARDELLA, PA
WALTER, SM
OKERHOLM, RA
机构
[1] Marion Merrell Dow Inc., Cincinnati, OH 45215-6300
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1991年 / 570卷 / 01期
关键词
D O I
10.1016/0378-4347(91)80208-T
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work describes the methodology for the analysis of terfenadine and the acid metabolite of terfenadine in plasma using high-performance liquid chromatography. The use of solid-phase extraction allows the use of robotic or manual sample preparation for the efficient clean-up of terfenadine and terfenadine acid metabolite from plasma. Additional selectivity is obtained through the use of fluorescence detection. For terfenadine, the validated quantitation range of this method is 10.0-84.2 ng/ml with coefficients of variation of 5.7-30%. For terfenadine acid metabolite, the validated quantitation range of this method is 8.2-500 ng/ml with coefficients of variation of 4.1-24%.
引用
收藏
页码:139 / 148
页数:10
相关论文
共 6 条
[1]  
ELLER MG, 1990, PHARM RES-DORDR, V7, pS206
[2]  
GARTEIZ DA, 1982, ARZNEIMITTEL-FORSCH, V32-2, P1185
[3]  
LEESON GA, 1982, ARZNEIMITTEL-FORSCH, V32-2, P1173
[4]   TERFENADINE - AN UPDATED REVIEW OF ITS PHARMACOLOGICAL PROPERTIES AND THERAPEUTIC EFFICACY [J].
MCTAVISH, D ;
GOA, KL ;
FERRILL, M .
DRUGS, 1990, 39 (04) :552-574
[5]   AN AUTOMATED-METHOD FOR THE DETERMINATION OF ENOXIMONE AND ITS MAJOR SULFOXIDE METABOLITE IN PLASMA USING ROBOTIC TECHNOLOGY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
WALTER, SM ;
CHAN, KY ;
COUTANT, JE .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1986, 9 (14) :3133-3155
[6]  
WOODWARD JK, 1988, CUTIS, V42, P5