LINE SHAPE ANALYSIS OF SPECTRA OBTAINED WITH GE(LI) DETECTORS

被引:11
作者
HUANG, FCP
OSMAN, CH
OPHEL, TR
机构
[1] Research School of Physical Sciences, Australian National University, Canberra
来源
NUCLEAR INSTRUMENTS & METHODS | 1969年 / 68卷 / 01期
关键词
D O I
10.1016/0029-554X(69)90701-0
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Line shape analysis techniques have been applied to gamma ray spectra obtained with a Ge(Li) detector. It was found that least-squares fitting procedures enabled more accurate and rapid extraction of relative intensities than was possible from the areas of the full energy or double escape peaks. © 1969.
引用
收藏
页码:141 / &
相关论文
共 6 条
  • [1] BRANCHING RATIOS OF 27AL BOUND STATES
    ELLIOTT, RV
    OPHEL, TR
    SPEAR, RH
    [J]. NUCLEAR PHYSICS A, 1968, A115 (03) : 673 - &
  • [2] FERGUSON AJ, 1962, AECL1398
  • [3] ACCURATE MEASUREMENT OF RELATIVE EFFICIENCY OF GE(LI) GAMMA-RAY DETECTORS IN ENERGY RANGE 500 TO 1500 KEV
    FREEMAN, JM
    JENKIN, JG
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1966, 43 (02): : 269 - &
  • [4] HUANG FCP, 1968, THESIS AUSTRALIAN NA
  • [5] OSMAN CH, TO BE PUBLISHED
  • [6] STORM E, 1958, LA2237 LOS AL SCI LA