X-RAY-MICROANALYSIS OF WET BIOLOGICAL SPECIMENS IN THE ENVIRONMENTAL SCANNING ELECTRON-MICROSCOPE .1. REDUCTION OF SPECIMEN DISTANCE UNDER DIFFERENT ATMOSPHERIC CONDITIONS

被引:23
作者
GILPIN, C
SIGEE, DC
机构
[1] UNIV MANCHESTER,SCH BIOL SCI,MANCHESTER M13 9PT,LANCS,ENGLAND
[2] UNIV MANCHESTER,SCH BIOL SCI,ELECTRON MICROSCOPE UNIT,MANCHESTER M13 9PT,LANCS,ENGLAND
来源
JOURNAL OF MICROSCOPY-OXFORD | 1995年 / 179卷
关键词
X-RAY MICROANALYSIS; ENVIRONMENTAL SCANNING ELECTRON MICROSCOPE; ELEMENTAL ANALYSIS; CYANOBACTERIA;
D O I
10.1111/j.1365-2818.1995.tb03610.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
X-ray microanalysis of non-biological and biological specimens was carried out in the environmental scanning electron microscope (ESEM) under different conditions of specimen distance (the distance travelled by the electron probe within the specimen chamber) and chamber atmosphere. Using both water vapour and argon atmospheres, it was shown that reduction in specimen distance had no effect on atmospheric gas X-ray signal in either case. Unlike water vapour, increased levels of argon (up to 10 torr) caused a marked depression of specimen P/B ratios, with a decrease in both characteristic and background (continuum) counts, These effects in argon were not altered by reduction in specimen distance. Specimen distance was important in relation to beam skirting and elemental analysis. With an extended assembly (short specimen distance), beam skirting in a water-vapour atmosphere was much reduced - leading to enhanced element detectability in a discrete biological specimen (Anabaena cyclindrica).
引用
收藏
页码:22 / 28
页数:7
相关论文
共 20 条
[1]  
ABRAHAM J, 1992, CELL BIOL INT, V17, P707
[2]   MINIMIZING SAMPLE EVAPORATION IN THE ENVIRONMENTAL SCANNING ELECTRON-MICROSCOPE [J].
CAMERON, RE ;
DONALD, AM .
JOURNAL OF MICROSCOPY-OXFORD, 1994, 173 :227-237
[3]  
CLAY S, 1991, SCANNING MICROSCOPY, V5, P207
[4]  
COLLINS SP, 1993, MICROSC RES TECHNIQ, V25, P406
[5]  
CZYZEWSKI Z, 1989, J MICROSC-OXFORD, V156, P285, DOI 10.1111/j.1365-2818.1989.tb02930.x
[6]  
DANILATOS GD, 1988, ADV ELECTRON EL PHYS, V71, P109
[7]   REVIEW AND OUTLINE OF ENVIRONMENTAL SEM AT PRESENT [J].
DANILATOS, GD .
JOURNAL OF MICROSCOPY, 1991, 162 :391-402
[8]   BIBLIOGRAPHY OF ENVIRONMENTAL SCANNING ELECTRON-MICROSCOPY [J].
DANILATOS, GD .
MICROSCOPY RESEARCH AND TECHNIQUE, 1993, 25 (5-6) :529-534
[9]  
DANILATOS GD, 1990, SCANNING MICROSCOPY, V4, P799
[10]   MECHANISMS OF DETECTION AND IMAGING IN THE ESEM [J].
DANILATOS, GD .
JOURNAL OF MICROSCOPY-OXFORD, 1990, 160 :9-19