Left Ventricular Assist Device Inflow Cannula Position May Contribute to the Development of HeartMate II Left Ventricular Assist Device Pump Thrombosis

被引:14
作者
Bhama, Jay K. [1 ]
Bansal, Aditya [2 ,3 ]
机构
[1] Univ Iowa Hlth Care, Dept Surg, 200 Hawkins Dr,SE500 GH, Iowa City, IA 52246 USA
[2] Ochsner Clin Fdn, Dept Surg, New Orleans, LA USA
[3] Univ Queensland, Sch Med, Ochsner Clin Sch, New Orleans, LA USA
关键词
Cannula; chest x-ray; thrombosis; ventricular assist device;
D O I
10.31486/toj.17.0070
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Pump thrombosis (PT) is a dreaded complication after left ventricular assist device (LVAD) implantation. Problems with inflow cannula (IC) position may precipitate thrombus development. We sought to determine if IC position contributes to the development of PT. Methods: We conducted a retrospective review of 76 HeartMate II LVAD implants. The angle of the IC (AIC) to the horizontal plane was measured on chest x-rays. Patients who developed PT (PT group) were compared to the remaining patients (control group). Results: The mean age at implantation was 56 +/- 14 years, and 82% of the patients were male. Ten patients (13%) developed PT. Six (60%) required device exchange, and 4 (40%) were managed with anticoagulation and/or thrombolysis. The median AIC for all patients at implantation was 59 degrees (range, 38 degrees-98 degrees; 25th-75th interquartile range, 50 degrees-75 degrees). In the PT group, the median AIC was larger at the time of PT diagnosis compared to implantation (70 degrees vs 60 degrees, P = 0.005). In the control group, the median AIC was also larger at follow-up compared to implantation (61 degrees vs 58 degrees, P < 0.001) although to a lesser degree than in the PT group. No difference was seen in the median AIC between the PT group and the control group at implantation (60 degrees vs 58 degrees, respectively; P = 0.668) or at follow-up (70 degrees vs 61 degrees, respectively; P = 0.309). However, the median AIC at follow-up in the PT group was significantly larger than the median AIC at implantation in the control group (70 degrees vs 58 degrees, respectively; P = 0.014). Conclusion: The HeartMate II LVAD IC position contributes to the development of PT. Regular monitoring of cannula position may help identify patients at risk for this problem.
引用
收藏
页码:131 / 135
页数:5
相关论文
共 9 条
[1]  
Bachman Timothy N, 2011, Cardiovasc Eng Technol, V2, P203
[2]   DEVELOPMENT OF AN EX VIVO OVINE VENTRICULAR ASSIST DEVICE MODEL FOR INTRAVENTRICULAR VISUALIZATION OF THE INFLOW CANNULA [J].
Bhama, Jay K. ;
Bachman, Timothy N. ;
Kormos, Robert L. ;
Borovetz, Harvey ;
Antaki, James F. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (08) :860-U5
[3]   Diaphragmatic implantation of the Heart Ware ventricular assist device [J].
Gregoric, Igor D. ;
Cohn, William E. ;
Frazier, O. H. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2011, 30 (04) :467-470
[4]   Presentation and management of left ventricular assist device inflow cannula malposition [J].
Milano, Carmelo A. ;
Simeone, Alan A. ;
Blue, Laura J. ;
Rogers, Joseph G. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2011, 30 (07) :838-840
[5]   Pump Replacement for Left Ventricular Assist Device Failure Can Be Done Safely and Is Associated With Low Mortality [J].
Moazami, Nader ;
Milano, Carmelo A. ;
John, Ranjit ;
Sun, Benjamin ;
Adamson, Robert M. ;
Pagani, Francis D. ;
Smedira, Nicholas ;
Slaughter, Mark S. ;
Farrar, David J. ;
Frazier, O. Howard .
ANNALS OF THORACIC SURGERY, 2013, 95 (02) :500-505
[6]   Performance of Extracorporeally Adjustable Ventricular Assist Device Inflow Cannula [J].
Shiose, Akira ;
Kim, Hyun-Il ;
Takaseya, Tohru ;
Kobayashi, Mariko ;
Fujiki, Masako ;
Chen, Ji-Feng ;
Klatte, Ryan ;
Gao, Shengqiang ;
Massiello, Alex L. ;
Saeed, Diyar ;
Fukamachi, Kiyotaka .
ANNALS OF THORACIC SURGERY, 2010, 90 (05) :1682-1687
[7]   Computed tomography correlates of inflow cannula malposition in a continuous-flow ventricular-assist device [J].
Sorensen, Erik N. ;
Hiivala, Nicholas J. ;
Jeudy, Jean ;
Rajagopal, Keshava ;
Griffith, Bartley P. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (06) :654-657
[8]   Device Exchange After Primary Left Ventricular Assist Device Implantation: Indications and Outcomes [J].
Stulak, John M. ;
Cowger, Jennifer ;
Haft, Jonathon W. ;
Romano, Matthew A. ;
Aaronson, Keith D. ;
Pagani, Francis D. .
ANNALS OF THORACIC SURGERY, 2013, 95 (04) :1262-1268
[9]   Surgical Technique Influences HeartMate II Left Ventricular Assist Device Thrombosis [J].
Taghavi, Sharven ;
Ward, Charisse ;
Jayarajan, Senthil N. ;
Gaughan, John ;
Wilson, Lynn M. ;
Mangi, Abeel A. .
ANNALS OF THORACIC SURGERY, 2013, 96 (04) :1259-1265