MODELS AND METHODS OF DETERMINING THE TRANSIENT-RESPONSE OF ION SENSORS - COMPARISON WITH IMPEDANCE SPECTROSCOPY

被引:2
|
作者
ATTARI, M
FABRY, P
QUEZEL, G
MALLIE, H
机构
[1] USTHB,INST ELECTR,DEPT INSTRUMENTAT,ALGIERS 16111,ALGERIA
[2] UNIV GRENOBLE 1,LAB INSTRUMENTAT MICROINFORMAT & ELECTR GRENOBLE,F-38000 GRENOBLE,FRANCE
关键词
ION SENSORS; IMPEDANCE SPECTROSCOPY; TRANSIENT RESPONSE;
D O I
10.1016/0925-4005(95)85163-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper deals with the study of models and methods for determining the transient response function of ion sensors (ion-selective electrodes, ISEs), especially in our case sodium sensors. Theoretical models have been studied, such as the Rechnitz-Hameka model, the Toth and Pungor model and the Markovic-Osborn model. The methods studied are the dipping method, which involves two separate solutions for step generation, and the injection method, which involves a carrier solution with a small sample solution, These methods are compared with a fast system realized for this purpose. Recorded data are fitted to mathematical models and the response time is determined and compared to the relaxation time obtained by impedance spectroscopy with a 1250-Schlumberger analyser. Actually, the results obtained show a very fast response time of about 50 ms for sensors based on NASICON material, which is ten times faster than those of usual sensors (Solea-Tacussel).