THE SUPER PET 3000-E - A PET SCANNER DESIGNED FOR HIGH COUNT RATE CARDIAC APPLICATIONS

被引:17
作者
TERPOGOSSIAN, MM
FICKE, DC
BEECHER, DE
HOFFMANN, GR
BERGMANN, SR
机构
[1] EDWARD MALLINCKRODT INST RADIOL,ELECTR RADIOL LAB,ST LOUIS,MO 63110
[2] WASHINGTON UNIV,SCH MED,DEPT INTERNAL MED,DIV CARDIOVASC,ST LOUIS,MO 63110
关键词
EMISSION COMPUTED TOMOGRAPHY; TECHNIQUES; APPARATUS AND EQUIPMENT; PHYSICS AND INSTRUMENTATION;
D O I
10.1097/00004728-199407000-00030
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Mathematical models for the delineation of regional myocardial perfusion and metabolism with PET require faithful reconstruction of arterial and myocardial time-activity curves following administration of radiotracers. High temporal resolution is often required in such measurements. Many commercially available tomographs exhibit long dead times that limit their count rate capabilities. To overcome these limitations, we developed and tested a whole-body tomographic device (Super PET 3000-E) with high count rate capabilities. The use of cesium fluoride scintillation detectors coupled with a one-to-one detector photomultiplier configuration reduces the system resolving and dead times. Materials and Methods: The Super PET 3000-E was subjected to a series of tests with phantoms to determine its resolution, sensitivity, linearity, count rate capabilities, dead time, and random coincidence contribution. Results: The system sensitivity is 136 kcounts/s/mu Ci/ml and its transverse and longitudinal resolutions are 8.5 and 10.5 mm full width at half-maximum, respectively. The system can easily record a total event rate of 2.0 Mcounts/s with minimal dead time loss and excellent linearity. Conclusion: The system fulfills its design goals and allows the very high count rate performance needed for the application of the physiological models used in our cardiac studies.
引用
收藏
页码:661 / 669
页数:9
相关论文
共 12 条
[1]   NONINVASIVE QUANTIFICATION OF REGIONAL MYOCARDIAL BLOOD-FLOW IN CORONARY-ARTERY DISEASE WITH OXYGEN-15-LABELED CARBON-DIOXIDE INHALATION AND POSITRON EMISSION TOMOGRAPHY [J].
ARAUJO, LI ;
LAMMERTSMA, AA ;
RHODES, CG ;
MCFALLS, EO ;
IIDA, H ;
RECHAVIA, E ;
GALASSI, A ;
DESILVA, R ;
JONES, T ;
MASERI, A .
CIRCULATION, 1991, 83 (03) :875-885
[2]   NONINVASIVE QUANTITATION OF MYOCARDIAL BLOOD-FLOW IN HUMAN-SUBJECTS WITH OXYGEN-15-LABELED WATER AND POSITRON EMISSION TOMOGRAPHY [J].
BERGMANN, SR ;
HERRERO, P ;
MARKHAM, J ;
WEINHEIMER, CJ ;
WALSH, MN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :639-652
[3]  
BERGMANN SR, 1992, POSITRON EMISSION TO, P97
[4]   EXTERNAL QUANTIFICATION OF MYOCARDIAL PERFUSION BY EXPONENTIAL INFUSION OF POSITRON-EMITTING RADIONUCLIDES [J].
HACK, SN ;
EICHLING, JO ;
BERGMANN, SR ;
WELCH, MJ ;
SOBEL, BE .
JOURNAL OF CLINICAL INVESTIGATION, 1980, 66 (05) :918-927
[5]   IMPLEMENTATION AND EVALUATION OF A 2-COMPARTMENT MODEL FOR QUANTIFICATION OF MYOCARDIAL PERFUSION WITH RB-82 AND POSITRON EMISSION TOMOGRAPHY [J].
HERRERO, P ;
MARKHAM, J ;
SHELTON, ME ;
BERGMANN, SR .
CIRCULATION RESEARCH, 1992, 70 (03) :496-507
[6]  
HOLMES TJ, 1984, IEE T MED IMAG, V2, P68
[7]  
SAMPATHKUMARAN K S, 1992, Journal of the American College of Cardiology, V19, p12A
[8]   A MATHEMATICAL-MODEL FOR POSITRON-EMISSION TOMOGRAPHY SYSTEMS HAVING TIME-OF-FLIGHT MEASUREMENTS [J].
SNYDER, DL ;
THOMAS, LJ ;
TERPOGOSSIAN, MM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (03) :3575-3583
[9]   MEASURING PET SCANNER SENSITIVITY - RELATING COUNTRATES TO IMAGE SIGNAL-TO-NOISE RATIOS USING NOISE EQUIVALENT COUNTS [J].
STROTHER, SC ;
CASEY, ME ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :783-788
[10]  
Ter-Pogossian M., 1992, POSITRON EMISSION TO, P1