REDUCTION OF POLYATOMIC INTERFERENCES IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY BY SELECTION OF INSTRUMENTAL PARAMETERS AND USING AN ARGON-NITROGEN PLASMA - EFFECT ON MULTIELEMENT ANALYSES

被引:35
作者
LABORDA, F
BAXTER, MJ
CREWS, HM
DENNIS, J
机构
关键词
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; POLYATOMIC INTERFERENCES; ARGON-NITROGEN PLASMA; MULTIELEMENT ANALYSIS;
D O I
10.1039/ja9940900727
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effect of instrumental parameters and argon-nitrogen plasmas on polyatomic ion formation has been studied in order to reduce their magnitude in routine multi-element analysis without losing detection capability. Special emphasis was placed on the chlorine based polyatomic interferences on V, Cr, Zn, As and Se. A significant reduction in signals from polyatomic ions was attained by using a high aerosol carrier gas flow rate (0.955 I min-1) instead of the default flow rate (0.755 I min-1), or by adding nitrogen (8%) to the aerosol carrier flow. The ArCl+ interference produced by 0.05% CI (the maximum concentration expected in digested food stuff samples) was effectively removed by both methods and CIO+ and CIO2+ by addition of nitrogen. Detection limits for elements along the mass range (from Li to U) were, on average, 2-3 times higher with the mixed gas plasma. This slight degradation of detection limits was not judged to be detrimental to multi-element determinations in five reference materials when the results from using an argon-only plasma (with low and high aerosol carrier flow rates) were compared with the results from the argon-nitrogen plasma.
引用
收藏
页码:727 / 736
页数:10
相关论文
共 28 条
[1]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[2]   INVESTIGATIONS ON MIXED-GAS PLASMAS PRODUCED BY ADDING NITROGEN TO THE PLASMA GAS IN ICP-MS [J].
BEAUCHEMIN, D ;
CRAIG, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (05) :603-614
[3]   DETERMINATION OF ARSENIC IN SAMPLES WITH HIGH CHLORIDE CONTENT BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BRANCH, S ;
EBDON, L ;
FORD, M ;
FOULKES, M ;
ONEILL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (02) :151-154
[4]   SAMPLE INTRODUCTION IN PLASMA EMISSION AND MASS-SPECTROMETRY - PLENARY LECTURE [J].
BROWNER, RF ;
ZHU, GX .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (06) :543-547
[5]   REDUCTION OF THE EFFECTS OF CONCOMITANT ELEMENTS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY BY ADDING NITROGEN TO THE PLASMA GAS [J].
CRAIG, JM ;
BEAUCHEMIN, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :937-942
[6]   INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - A REVIEW [J].
EVANS, EH ;
GIGLIO, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :1-18
[7]   SIMPLE APPROACH TO REDUCING POLYATOMIC ION INTERFERENCES ON ARSENIC AND SELENIUM IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (03) :299-300
[8]   EFFECT OF ORGANIC-SOLVENTS AND MOLECULAR GASES ON POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
EVANS, EH ;
EBDON, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (06) :425-429
[9]  
FORD M, 1992, ANAL PROC, V29, P104