STEAM STERILIZATION OF LAPAROSCOPIC INSTRUMENTS

被引:0
作者
VOYLES, CR [1 ]
SANDERS, DL [1 ]
SIMONS, JE [1 ]
MCVEY, EA [1 ]
WILSON, WB [1 ]
机构
[1] UNIV MISSISSIPPI,SCH MED,JACKSON,MS 39216
关键词
STERILIZATION; CONTAMINATION; LAPAROSCOPY;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
Because of the intricate internal parts of laparoscopic instruments, questions have been raised about the efficacy of cleaning and sterilization techniques. To assess these risks, hamburger meat was inoculated with high concentrations of vegetative pathogens and packed into laparoscopic cannulas. All openings of the cannulas were sealed during steam sterilization cycles ranging from 3 to 10 min in different experiments; cultures were obtained after cooling. Experiments were then performed using heat-resistant spore forms. Our studies showed that both the standard 10-min cycle and the 3-min ''flash'' were uniformly successful in killing all pathogenic microorganisms. A 7-min steam sterilization was necessary to kill spores within sealed cannulas. We conclude that a standard 10-min cycle within the steam autoclave provides complete sterilization of laparoscopic instruments; the 3-min ''flash'' sterilization is appropriate and safe for instruments that have been inadvertently contaminated or dropped during a surgical procedure.
引用
收藏
页码:139 / 141
页数:3
相关论文
共 10 条
[1]  
BIGELOW WD, 1921, J INFECT DIS, V20, P528
[2]  
Cahill N, 1993, Bull Am Coll Surg, V78, P28
[3]  
CORSON SL, 1979, J REPROD MED, V23, P49
[4]  
Diettrich N A, 1991, J Laparoendosc Surg, V1, P295, DOI 10.1089/lps.1991.1.295
[5]  
Harris M H, 1992, AORN J, V55, P1547, DOI 10.1016/S0001-2092(07)66592-2
[6]  
MAKI DG, 1977, J SURG PRACTICE, V5, P10
[7]  
PERKINS JJ, 1978, PRINCIPLES METHODS S, P63
[8]  
SERWAY RA, 1990, PHYSICS SCI ENG, P533
[9]  
VOYLES CR, 1993, AM J SURG, V65, P472
[10]  
1992, AORN J, V55, P228