RECENT RESULTS IN SCINTILLATION DETECTION WITH SILICON AVALANCHE PHOTODIODES

被引:23
作者
CARRIER, C
LECOMTE, R
机构
[1] Department of Nuclear Medicine and Radiobiology, University of Sherbrooke, Sherbrooke
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1109/23.106620
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We report on the investigation of silicon avalanche photodiodes in scintillation detection for potential applications in medical imaging and nuclear spectroscopy. Experimental measurements were performed on BGO, CdWO4, plastic, NaI(T1) and CsI(T1) scintillators, geometrically optimized for light collection. Results show the advantages of a solid state photodetector having an internal gain for the detection of low energy gamma and X-rays. When coupled to CsI(T1), X-ray energies as low as 5.9 keV can be discriminated from the noise. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:209 / 214
页数:6
相关论文
共 8 条
[1]  
GRASMAN H, 1985, NUCL INSTR METH PHYS, V228, P313
[3]   A MEASUREMENT OF THE LIGHT YIELD OF COMMON INORGANIC SCINTILLATORS [J].
HOLL, I ;
LORENZ, E ;
MAGERAS, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :105-109
[4]  
KNOLL GF, 1979, RAD DETECTION MEASUR, P254
[5]   STATUS OF BGO-AVALANCHE PHOTODIODE DETECTORS FOR SPECTROSCOPIC AND TIMING MEASUREMENTS [J].
LECOMTE, R ;
MARTEL, C ;
CARRIER, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 278 (02) :585-597
[6]  
LECOMTE R, 1985, IEEE T NUCL SCI, V32, P482, DOI 10.1109/TNS.1985.4336878
[7]   ULTRA-LOW NOISE CHARGE SENSITIVE PREAMPLIFIER FOR SCINTILLATION DETECTION WITH AVALANCHE PHOTODIODES IN PET APPLICATIONS [J].
SCHMITT, D ;
LECOMTE, R ;
LAPOINTE, M ;
MARTEL, C ;
CARRIER, C ;
KARUTA, B ;
DUVAL, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :91-96
[8]  
WEBB PP, 1974, RCA REV, V35, P234