DETERMINATION OF THE CYSTEINE DERIVATIVES N-ACETYLCYSTEINE, S-CARBOXYMETHYLCYSTEINE AND METHYLCYSTEINE IN PHARMACEUTICALS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY

被引:24
作者
TSAI, FY
CHEN, CJ
CHIEN, CS
机构
[1] Department of Health, National Laboratories of Foods and Drugs, Nankang, Taipei
关键词
D O I
10.1016/0021-9673(94)00959-D
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A high-performance liquid chromatographic method was developed for the determination of three mucolytic agents, N-acetylcysteine (NAGS), S-carboxymethylcysteine (SCMCS) and methylcysteine (MCS). Chromatography was performed on an ODS column using an isocratic mobile phase consisting of acetonitrile, methanol and sodium hexanesulfonic acid buffer (pH 2.9) at a flow-rate of 1.0 ml/min, with UV detection at 220 nm. Methionine was used as an internal standard. Good quantitative results were obtained, the recoveries from synthetic mixtures being >98.2% with R.S.D. <1%. Linearity over a concentration range of over one order magnitude was obtained; the correlation coefficients were >0.9997 at concentrations of 0.05, 0.1, 1.0, 2.0 and 3.0 mg/ml. The addition of methanol and acetonitrile to the mobile phase improved the peak shape and retention of the compounds of interest. The selectivity of the method was demonstrated by separating four similar thiol compounds and determination of the analytes in pharmaceuticals.
引用
收藏
页码:309 / 315
页数:7
相关论文
共 18 条
[1]  
Bovis G, 1980, Boll Chim Farm, V119, P172
[2]   QUANTITATIVE-DETERMINATION OF GLUTATHIONE, CYSTEINE, AND N-ACETYL CYSTEINE CONJUGATES OF ACETAMINOPHEN BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
BUCKPITT, AR ;
ROLLINS, DE ;
NELSON, SD ;
FRANKLIN, RB ;
MITCHELL, JR .
ANALYTICAL BIOCHEMISTRY, 1977, 83 (01) :168-177
[3]   USE OF METHIONINE FOR ACUTE PARACETAMOL POISONING [J].
CROME, P ;
VOLANS, GN ;
VALE, JA ;
WIDDOP, B ;
GOULDING, R ;
WILLIAMS, RS .
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 1976, 4 :105-111
[4]  
GOODMANN LA, 1988, PHARMACOL BASIS THER, P702
[5]   DETERMINATION OF 3-MERCAPTOLACTATE, MERCAPTOACETATE AND N-ACETYLCYSTEINE IN URINE BY GAS-CHROMATOGRAPHY [J].
HANNESTAD, U ;
SORBO, B .
CLINICA CHIMICA ACTA, 1979, 95 (02) :189-200
[6]   DETERMINATION OF CYSTEINE, GLUTATHIONE AND N-ACETYLCYSTEINE IN PLASMA BY ION-PAIR REVERSED-PHASE LIQUID-CHROMATOGRAPHY AND POST-COLUMN DERIVATIZATION [J].
JOHANSSON, M ;
LENNGREN, S .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 432 :65-74
[7]  
KENICHIRO N, 1987, J CHROMATOGR, V414, P11
[8]   IMPROVED METHOD FOR THE DETERMINATION OF N-ACETYLCYSTEINE IN HUMAN-PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
LEWIS, PA ;
WOODWARD, AJ ;
MADDOCK, J .
JOURNAL OF CHROMATOGRAPHY, 1985, 327 (JUN) :261-267
[9]   APPROPRIATE USE OF MUCOLYTIC AGENTS [J].
LIEBERMA.J .
AMERICAN JOURNAL OF MEDICINE, 1970, 49 (01) :1-&
[10]   DETERMINATION OF S-CARBOXYMETHYLCYSTEINE IN SYRUP FORMULATIONS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
MELUCCI, CK ;
LYMAN, GW ;
BOND, AD ;
JOHNSON, RN .
JOURNAL OF CHROMATOGRAPHY, 1987, 391 (01) :321-324