MEASUREMENT OF VOLATILE ORGANICS AT PART PER BILLION CONCENTRATIONS USING A COLD TRAP INLET AND HIGH-SPEED GAS-CHROMATOGRAPHY

被引:9
作者
MOURADIAN, RF [1 ]
LEVINE, SP [1 ]
KE, HQ [1 ]
ALVORD, HH [1 ]
机构
[1] UNIV MICHIGAN,SCH PUBL HLTH,DEPT ENVIRONM & IND HLTH,ANN ARBOR,MI 48109
来源
JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION | 1991年 / 41卷 / 08期
关键词
D O I
10.1080/10473289.1991.10466901
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Direct inlet gas chromatography (GC) is becoming one of the most frequently used techniques for measurement of volatile organics in air. Although GC is an effective tool, its usefulness as a field method can be limited by retention times of several minutes or more, and by the limits of detection. In order to address these issues, a high speed GC system featuring a capillary cold trap inlet and variable speed electrometer-amplifier was developed and tested. The gas cooled inlet was used to collect and focus organic vapors from injection volumes of up to 1 ml. Resistance heating of the metal cold trap produced a narrow injection band that allowed simple separations to be completed in 5 to 20 seconds. Use of a variable speed electrometer-amplifier allowed the response time and noise suppression characteristics to be adjusted to the needs of the specific analysis. Simple mixtures of organics including aromatics, alkanes and chlorinated hydrocarbons were separated in 20 seconds or less. The limits of detection for 13 test compounds ranged from less than 1 part per billion by volume (ppbv) to about 50 ppbv. Data presented here indicate that high speed GC using a cold trap inlet may be a feasible approach for near real-time measurement of volatile organics in ambient air.
引用
收藏
页码:1067 / 1072
页数:6
相关论文
共 30 条
[1]   THE USE OF COANDA WALL ATTACHMENT FLUIDIC SWITCHES AS GAS-CHROMATOGRAPHIC VALVES [J].
ANNINO, R ;
LEONE, J .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1982, 20 (01) :19-26
[2]  
DESTY DH, 1965, ADV CHROMATOGR, V1, P199
[3]   INFLUENCE OF INSTRUMENTAL CONTRIBUTIONS ON APPARENT COLUMN EFFICIENCY IN HIGH-SPEED GAS-CHROMATOGRAPHY [J].
GASPAR, G ;
ANNINO, R ;
VIDALMADJAR, C ;
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1978, 50 (11) :1512-1518
[4]   STUDY IN HIGH-SPEED GAS-CHROMATOGRAPHY .1. INJECTIONS OF NARROW SAMPLE PLUGS [J].
GASPAR, G ;
ARPINO, P ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1977, 15 (07) :256-261
[5]   EVALUATION OF CANISTERS FOR MEASURING EMISSIONS OF VOLATILE ORGANIC AIR-POLLUTANTS FROM HAZARDOUS-WASTE INCINERATION [J].
GHOLSON, AR ;
STORM, JF ;
JAYANTY, RKM ;
FUERST, RG ;
LOGAN, TJ ;
MIDGETT, MR .
JAPCA-THE JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 1989, 39 (09) :1210-1217
[6]   SPEED OF ANALYSIS IN OPEN TUBULAR COLUMNS GAS-CHROMATOGRAPHY [J].
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1978, 50 (13) :1812-1821
[7]   AMBIENT CAPILLARY CHROMATOGRAPHY OF VOLATILE ORGANICS WITH A PORTABLE GAS-CHROMATOGRAPH [J].
JERPE, J ;
DAVIS, A .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1987, 25 (04) :154-157
[8]   IMPROVEMENT OF SPEED OF SEPARATION IN PACKED-COLUMN GAS-CHROMATOGRAPHY [J].
JONKER, RJ ;
POPPE, H ;
HUBER, JFK .
ANALYTICAL CHEMISTRY, 1982, 54 (14) :2447-2456
[9]   COLLECTION AND ANALYSIS OF HAZARDOUS ORGANIC EMISSIONS [J].
KROST, KJ ;
PELLIZZARI, ED ;
WALBURN, SG ;
HUBBARD, SA .
ANALYTICAL CHEMISTRY, 1982, 54 (04) :810-817
[10]   ELECTRICALLY HEATED COLD TRAP INLET SYSTEM FOR COMPUTER-CONTROLLED HIGH-SPEED GAS-CHROMATOGRAPHY [J].
LANNING, LA ;
SACKS, RD ;
MOURADIAN, RF ;
LEVINE, SP ;
FOULKE, JA .
ANALYTICAL CHEMISTRY, 1988, 60 (18) :1994-1996