AN ANTISCATTER DEVICE FOR LOW-TEMPERATURE CRYSTALLOGRAPHIC EXPERIMENTS WITH IMAGING PLATES

被引:23
作者
DAROVSKY, A [1 ]
BOLOTOVSKY, R [1 ]
COPPENS, P [1 ]
机构
[1] SUNY BUFFALO,DEPT CHEM,BUFFALO,NY 14260
关键词
D O I
10.1107/S0021889894006369
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The use of area detectors in low-temperature crystallographic data collection requires new methods for elimination of scattering by cryostat chambers, as diffracted-beam collimators can no longer be employed. The antiscatter device described is mounted inside the vacuum chamber and kept stationary during sample rotation by the use of a magnetic clamp. The device has been tested on both conventional and synchrotron sources.
引用
收藏
页码:1039 / 1040
页数:2
相关论文
共 2 条
[1]   CLOSED-CYCLE HELIUM CRYOSTAT MOUNT FOR THE HUBER 511.1 DIFFRACTOMETER CIRCLE [J].
GRAAFSMA, H ;
SAGERMAN, G ;
COPPENS, P .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1991, 24 :961-962
[2]   MOUNTING A 10-K COOLING DEVICE WITHOUT ROTATING SEALS ON A 4-CIRCLE DIFFRACTOMETER [J].
HENRIKSEN, K ;
LARSEN, FK ;
RASMUSSEN, SE .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1986, 19 :390-394