The effect of X-ray beam distortion on the Edwards Sapien XT (TM) trans-catheter aortic valve replacement prosthesis

被引:1
作者
Crowhurst, James A. [1 ,2 ,3 ]
Poon, Karl K. [1 ]
Murdoch, Dale [1 ,2 ]
Incani, Alexander [1 ]
Raffel, Owen C. [1 ,2 ]
Liddicoat, Annelise [3 ]
Walters, Darren [1 ,2 ]
机构
[1] Prince Charles Hosp, Heart & Lung Inst, Brisbane, Qld 4032, Australia
[2] Univ Queensland, Brisbane, Qld, Australia
[3] Prince Charles Hosp, Dept Med Imaging, Brisbane, Qld 4032, Australia
关键词
C-arm rotation; distortion; fluoroscopy; percutaneous; TAVR; X-ray;
D O I
10.1002/jmrs.131
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Profiling the Aortic root perpendicular to the fluoroscopic image plane will achieve a more successful implant position for trans-catheter aortic valve replacement (TAVR). This study aimed to investigate whether the divergent nature of the X-ray beam from the C-arm altered the appearance of the TAVR device. Methods: Under bench-top testing, a 23, 26 and 29 mm Edwards Sapien XT valve was positioned coaxially at the bottom of a fluoroscopic image utilising 22 and 32 cm fields of view (FOV). The table was then moved so that the valve was positioned at the top of the image. The valve's appearance was scored using a previously published three tier classification tool (excellent, satisfactory and poor) and quantified with measurements. The number of degrees of C-arm rotation that were required to bring the valve back to a coaxial appearance was recorded. Results: When using the 32 cm FOV, the valve's appearance changes from excellent to satisfactory. When a 22 cm FOV was used, the change is less marked. More C-arm rotation is required to bring the appearance back to coaxial with the 32 cm FOV. Conclusion: Not maintaining the valve in the centre of the image can distort the valves appearance. This has the potential to affect the final implantation depth.
引用
收藏
页码:239 / 245
页数:7
相关论文
共 17 条
[1]   Prevention and Management of Transcatheter Balloon-Expandable Aortic Valve Malposition [J].
Al Ali, Abdullah M. ;
Altwegg, Lukas ;
Horlick, Eric M. ;
Feindel, Christopher ;
Thompson, Christopher R. ;
Cheung, Anson ;
Carere, Ronald G. ;
Humphries, Karin ;
Ye, Jian ;
Masson, Jean-Bernard ;
Webb, John G. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 72 (04) :573-578
[2]   Transaortic Transcatheter Aortic Valve Implantation: Step-by-Step Guide [J].
Bapat, Vinnie ;
Attia, Rizwan .
SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2012, 24 (03) :206-211
[3]   Impact of Post-Implant SAPIEN XT Geometry and Position on Conduction Disturbances, Hemodynamic Performance, and Paravalvular Regurgitation [J].
Binder, Ronald K. ;
Webb, John G. ;
Toggweiler, Stefan ;
Freeman, Melanie ;
Barbanti, Marco ;
Willson, Alexander B. ;
Alhassan, Donya ;
Hague, Cameron J. ;
Wood, David A. ;
Leipsic, Jonathon .
JACC-CARDIOVASCULAR INTERVENTIONS, 2013, 6 (05) :462-468
[4]   Prediction of Optimal Deployment Projection for Transcatheter Aortic Valve Replacement Angiographic 3-Dimensional Reconstruction of the Aortic Root Versus Multidetector Computed Tomography [J].
Binder, Ronald K. ;
Leipsic, Jonathon ;
Wood, David ;
Moore, Teri ;
Toggweiler, Stefan ;
Willson, Alex ;
Gurvitch, Ronen ;
Freeman, Melanie ;
Webb, John G. .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2012, 5 (02) :247-252
[5]  
Bontrager KL, 2005, TXB RADIOGRAPHIC POS, V6th
[6]  
Buschong SC., 1993, RADIOLOGIC SCI TECHN
[7]   Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis [J].
Cribier, A ;
Eltchaninoff, H ;
Tron, C ;
Bauer, F ;
Agatiello, C ;
Sebagh, L ;
Bash, A ;
Nusimovici, D ;
Litzler, PY ;
Bessou, JP ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (04) :698-703
[8]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[9]   Precision measurement of disc height, vertebral height and sagittal plane displacement from lateral radiographic views of the lumbar spine [J].
Frobin, W ;
Brinckmann, P ;
Biggemann, M ;
Tillotson, M ;
Burton, K .
CLINICAL BIOMECHANICS, 1997, 12 :S1-S63
[10]   Multislice Computed Tomography for Prediction of Optimal Angiographic Deployment Projections During Transcatheter Aortic Valve Implantation [J].
Gurvitch, Ronen ;
Wood, David A. ;
Leipsic, Jonathon ;
Tay, Edgar ;
Johnson, Mark ;
Ye, Jian ;
Nietlispach, Fabian ;
Wijesinghe, Namal ;
Cheung, Anson ;
Webb, John G. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (11) :1157-1165