PERFORMANCE OF A SINGLE-CRYSTAL DIGITAL GAMMA-CAMERA FOR 1ST PASS CARDIAC STUDIES

被引:3
作者
DOBBELEIR, A
FRANKEN, PR
HAM, HR
BRIHAYE, C
GUILLAUME, M
KNAPP, FF
VANDEVIVERE, J
机构
[1] Division of Nuclear Medicine, Middelheim General Hospital, Antwerpen
[2] Division of Nuclear Medicine, St Peter's Hospital, Bruxelles
[3] Cyclotron Research Center, University of Liège, Liège
[4] Nuclear Medicine Group, Health and Safety Research Division, Oak Ridge National Laboratory (ORNL), Oak Ridge
关键词
D O I
10.1097/00006231-199101000-00004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
First pass radionuclide angiocardiography (FPRNA) has gained increasing interest because of the development of new Tc-99m-labelled perfusion agents and of new Os-191/Ir-191m generator systems. The aim of the study was to evaluate the performance capacities of a small field of view single crystal digital gamma camera for Tc-99m and Ir-191m at high count rates. The camera dead time for Tc-99m (window 30%) was well corrected up to 300 kcps in fast acquisition mode using the relative decrease of a small shielded reference source. Using the decaying activity method for Ir-191m the non-linearity response of the gamma camera was corrected by an Os-191 reference source up to 210 kcps at 70 keV, 75 kcps at 129 keV and 320 kcps including both peaks. Saturation count rates were respectively 270 kcps, 150 kcps and 420 kcps and high count rate resolution (FWHM) 9.0, 7.3 and 10.3 mm. Since the accuracy of first pass measurements is more sensitive to count rate than to spatial resolution the 50-150 keV window was chosen for clinical studies. In data obtained from 32 ECG gated FPRNA patient studies, the whole field of view count rate during the left ventricular phase ranged from 100 to 250 kcps with 80 to 120 mCi (2960-4400 MBq) of Ir-191m and 100 to 180 kcps with 20 to 25 mCi (750-925 MBq) of Tc-99m red blood cells permitting for both tracers accurate non-linearity correction.
引用
收藏
页码:27 / 34
页数:8
相关论文
共 14 条
[1]  
Sporn V., Perez Balino N., Holman B.L., Et al., Simultaneous measurement of ventricular function and myocardial perfusion using the technetium-99m isonitriles, Clin Nucl Med, 13, pp. 77-81, (1988)
[2]  
Baillet G.Y., Mena I.G., Kuperus J.H., Et al., Simultaneous technetium-99m MIBI angiography and myocardial perfusion imaging, J Nucl Med, 30, pp. 38-44, (1989)
[3]  
Brihaye C., Butler T.A., Knapp F.F., Et al., A new osmium-191/iridium-191m radionuclide generator system using activated carbon, J Nucl Med, 27, pp. 380-387, (1986)
[4]  
Packard A.B., Treves S.T., O'brien G.M., Lim K.S., An osmium-191/iridium-191m radionuclide generator using an oxalato osmate parent complex, J Nucl Med, 28, pp. 1571-1576, (1987)
[5]  
Issachar D., Abrashkins S., Weiniger J., Et al., Osmium-191/iridium-191m generator based on silica gel impregnated with tridodecylmethylammonium chloride, J Nucl Med, 30, pp. 538-541, (1989)
[6]  
Heller G.V., Treves S.T., Parker J.A., Et al., Comparison of ultrashort-lived iridium-191m with technetium-99m for first pass radionuclide angiocardiographic evaluation of right and left ventricular function in adults, J am Coll Cardiol, 7, pp. 1295-1302, (1986)
[7]  
Heilman C., Zafrir N., Shimoni A., Et al., Evaluation of ventricular function with first pass iridium-191m radionuclide angiocardiography, J Nucl Med, 30, pp. 450-457, (1989)
[8]  
Franken P.R., Dobbeleir A., Ham H.R., Et al., Clinical usefulness of ultrashort-lived iridium- 191m from a carbon-based generator system for the evaluation of the left ventricular function, J Nucl Med, 30, pp. 1025-1031, (1989)
[9]  
Brihaye C., Dewez S., Guillaume M., Et al., Reactor production and purification of osmium-191 for use in a new Os-191/Ir-191m radionuclide generator system, Appl Radiat Isot, 40, pp. 183-189, (1989)
[10]  
Ullman V., Husak V., Dubroka L., Deadtime correction in dynamic radionuclides studies by computer, Eur J Nucl Med, 3, pp. 197-202, (1978)