ONLINE ELECTROLYTIC DISSOLUTION OF ALLOYS AND MULTIELEMENT DETERMINATION BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY

被引:22
|
作者
YUAN, DX
WANG, XR
YANG, PY
HUANG, BL
机构
[1] Department of Chemistry, Xiamen University, Xiamen
关键词
EMISSION SPECTROMETRY; FLOW SYSTEM; ALUMINUM ALLOYS; ELECTROLYTIC DISSOLUTION; PLASMAS;
D O I
10.1016/0003-2670(91)87134-S
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the direct simultaneous multi-element analysis of solid metal samples by flow-injection on-line electrolytic dissolution and inductively coupled plasma atomic emission spectrometry was developed and applied to the determination of Zn, Si, Fe, Mn, Cr, Mg and Cu in aluminium alloys. The suitable range for the determination was 10-mu-g g-1-100 mg g-1. The precision of the entire analysis, including electrolysis and determination, was in the range 1-6%, depending on the element and its concentration in the alloys and on the current density applied in electrolysis. The analysis of a sample can be accomplished in a few minutes.