VIOLATION OF THE GRADIENT APPROXIMATION IN NMR SELF-DIFFUSION MEASUREMENTS

被引:7
作者
STEPISNIK, J [1 ]
机构
[1] JOZEF STEFAN INST, LJUBLJANA 61111, SLOVENIA
来源
ZEITSCHRIFT FUR PHYSIKALISCHE CHEMIE-INTERNATIONAL JOURNAL OF RESEARCH IN PHYSICAL CHEMISTRY & CHEMICAL PHYSICS | 1995年 / 190卷
关键词
NMR SELF-DIFFUSION; MEASUREMENT FIELD GRADIENT METHOD; INHOMOGENEOUS MAGNETIC FIELD; FRINGE FIELD; GEOMAGNETIC FIELD;
D O I
10.1524/zpch.1995.190.Part_1.051
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The gradient approximation of magnetic field fails at NMR self-diffusion measurement with a very strong nonuniform magnetic field. The standard formula for self-diffusion attenuation of the spin-echoes has to be replaced by a new one where the role of the gadient of one field component ist taken over by the gradient of the magnitude of the total magnetic field. It leads to a distribution of the spin-echo attenuation in specimens, and to a nonexponential relation between the spin-echo signal and parameters of the applied nonuniform magnetic field. The anisotropy of particle migrations furthermore enlarges the nonuniformity of the attenuation distribution.
引用
收藏
页码:51 / 62
页数:12
相关论文
共 19 条