SALINE-FILLED CUFFS HELP PREVENT LASER-INDUCED POLYVINYLCHLORIDE ENDOTRACHEAL-TUBE FIRES

被引:17
作者
SOSIS, MB [1 ]
DILLON, FX [1 ]
机构
[1] INDIANA UNIV,SCH MED,DEPT ANESTHESIA,INDIANAPOLIS,IN 46202
关键词
EQUIPMENT; CUFFS ENDOTRACHEAL; LASERS CUFF FLAMMABILITY;
D O I
10.1213/00000539-199102000-00008
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
To determine whether the filling of tracheal tube cuffs with saline would decrease their combustibility during laser surgery, 20 polyvinylchloride tracheal tubes were studied. The cuffed end of each tracheal tube was inserted into the neck of an empty flask, and the tube and flask were flushed with oxygen for 5 min before cuff inflation. Ten tracheal tubes had their cuffs inflated with air, and 10 were inflated with saline. A Lasersonics LS880 CO2 laser, set to 5 W for five of each of the two types of filled cuffs and to 40 W for the other pair of five tubes, was fired continuously at the cuffs for up to 1 min. No combustion occurred at the 5-W setting. The times to cuff perforation when the laser was set at 5 W were (mean +/- SD) 1.00 +/- 0.83 and 4.21 +/- 3.91 s for the air- and saline-filled cuffs, respectively, a difference that was not statistically significant. The time to deflation of the saline-filled cuff (104.6 +/- 67.5 s) was, however, significantly longer than that of the air-filled cuff (2.59 +/- 1.97 s). When the tracheal tube cuffs were exposed to 40-W laser radiation, the cuff and adjacent tube shaft ignited in all cases when the cuffs were inflated with air, but only in one of five cases when the cuffs were filled with saline (P < 0.05). The filling of tracheal tube cuffs with saline provides simple, moderately effective partial protection of the cuff of endotracheal tubes during CO2 laser airway surgery.
引用
收藏
页码:187 / 189
页数:3
相关论文
共 7 条
[1]   LASER-INDUCED ENDOTRACHEAL-TUBE FIRE [J].
COZINE, K ;
ROSENBAUM, LM ;
ASKANAZI, J ;
ROSENBAUM, SH .
ANESTHESIOLOGY, 1981, 55 (05) :583-585
[2]  
FRIED MP, 1983, LARYNGOSCOPE, V93, P275
[3]   HEAT SINK PROTECTION AGAINST LASERING ENDOTRACHEAL CUFFS [J].
LEJEUNE, FE ;
LETARD, F ;
GUICE, C ;
MARICE, H .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1982, 91 (06) :606-607
[4]   HAZARD TO ENDOTRACHEAL-TUBES BY CO2-LASER BEAM - EXPERIMENTAL REPORT [J].
OHASHI, N ;
ASAI, M ;
UEDA, S ;
IMAMURA, J ;
WATANABE, Y ;
MIZUKOSHI, K .
ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 1985, 47 (01) :22-25
[5]  
SNOW JC, 1976, ANESTH ANALG, V55, P146
[6]  
SOSIS M, 1990, ANESTHESIOLOGY, V74, P747
[7]  
SOSIS MB, 1989, ANESTH ANALG, V68, P392