MEASUREMENT OF TRABECULAR BONE-MINERAL DENSITY IN THE DISTAL RADIUS BY 2 GAMMA-RAY COMPUTED-TOMOGRAPHY SCANNERS

被引:9
作者
HOSIE, CJ [1 ]
机构
[1] W SCOTLAND HLTH BOARDS,DEPT CLIN PHYS & BIOENGN,GLASGOW G4 9LF,SCOTLAND
关键词
D O I
10.1088/0967-3334/14/3/005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The precision and measurement agreement of two gamma-ray computed tomography forearm bone scanners have been evaluated. The inherent precision in vitro for the Stratec SCT 900 scanner was 0.28% compared with 0.04% for the OSCAR scanner. The measurements were linear over the range of densities found in humans. Increasing thickness of the cortical shell dramatically increased trabecular bone attenuation coefficient (TBAC) for the SCT 900, but only slightly affected results with the OSCAR. Short-term precision of the SCT 900 in 22 women (1.26%) was not as good as that previously reported for the OSCAR (0.50%). The measured TBAC for the same patients was significantly greater with the SCT 900. Corrections for the error introduced by the cortical shell of the radius with the SCT were calculated. Using this correction similar TBAC values could be obtained with both scanners, but the variance of the differences was too great for the results to be used interchangeably. Published bone density data for normal subjects indicate that Scottish women lose more trabecular bone in the two decades after 50 years of age than those in the Federal Republic of Germany.
引用
收藏
页码:269 / 276
页数:8
相关论文
共 14 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   CORRECTION OF SCATTER IN COMPUTED-TOMOGRAPHY IMAGES OF BONE [J].
HANGARTNER, TN .
MEDICAL PHYSICS, 1987, 14 (03) :335-340
[3]  
HANGARTNER TN, 1982, J COMPUT ASSIST TOMO, V6, P11560
[4]  
HESP R, 1985, CLIN PHYS PHYSL MEAS, V4, P303
[5]   A GAMMA-RAY COMPUTED-TOMOGRAPHY SCANNER FOR THE QUANTITATIVE MEASUREMENT OF BONE-DENSITY [J].
HOSIE, CJ ;
RICHARDSON, W ;
GREGORY, NL .
JOURNAL OF BIOMEDICAL ENGINEERING, 1985, 7 (01) :30-34
[6]   DIFFERENTIAL EFFECT OF LONG-TERM ESTROGEN THERAPY ON TRABECULAR AND CORTICAL BONE [J].
HOSIE, CJ ;
HART, DM ;
SMITH, DAS ;
ALAZZAWI, F .
MATURITAS, 1989, 11 (02) :137-145
[7]   PRECISION OF MEASUREMENT OF BONE-DENSITY WITH A SPECIAL PURPOSE COMPUTED-TOMOGRAPHY SCANNER [J].
HOSIE, CJ ;
SMITH, DAS .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (700) :345-350
[8]  
HOSIE CJ, 1986, THESIS COUNCIL NATIO
[9]   THE EXPONENTIAL EDGE-GRADIENT EFFECT IN X-RAY COMPUTED-TOMOGRAPHY [J].
JOSEPH, PM ;
SPITAL, RD .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :473-487
[10]  
RUEGSEGGER P, 1981, J COMPUT ASSIST TOMO, V5, P384