AUTOMATIC BRIDGE FOR INDUCTIVE VOLTAGE DIVIDERS

被引:4
作者
CHANG, P
LIANG, CP
机构
[1] Electrical Measurement Laboratory, Center for Measurement Standards, I.T.R.I., Hsinchu
关键词
D O I
10.1109/19.377868
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We describe an automatic ac voltage ratio standard based on an injection-type bridge and two test inductive voltage dividers (IVD) for applications at low audio frequencies. The ratio standard is generated by self-calibrating two test IVD's by an automated ''boot-strap'' procedure. The bridge is the heart of our reference standard for ac voltage ratios as well as a calibration system for IVD's. The ratio accuracy of our bridge calibration is better than 0.05 ppm at 500 Hz. In addition, we describe two IVD designs used in this work.
引用
收藏
页码:418 / 421
页数:4
相关论文
共 4 条
[1]   AUTOMATIC INDUCTIVE VOLTAGE DIVIDER BRIDGE FOR OPERATION FROM 10 HZ TO 100 KHZ [J].
AVRAMOV, S ;
OLDHAM, NM ;
JARRETT, DG ;
WALTRIP, BC .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1993, 42 (02) :131-135
[3]   AN INJECTION METHOD FOR SELF-CALIBRATION OF INDUCTIVE VOLTAGE DIVIDERS [J].
SZE, WC .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION C-ENGINEERING AND INSTRUMENTATION, 1968, C 72 (01) :49-+