MEASUREMENTS AND CALCULATIONS OF THE ABSORBED DOSE DISTRIBUTION AROUND A CO-60 SOURCE

被引:7
作者
TIOURINA, TB
DRIES, WJF
VANDERLINDEN, PM
机构
[1] Department of Radiotherapy, Eindhoven, Catharina Hospital
关键词
BRACHYTHERAPY; DOSE RATIO FUNCTION; MONTE CARLO CODE;
D O I
10.1118/1.597583
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The data from Meisberger et al. [Radiology 90, 953–957 (1968)] are often used as a basis for dose calculations in brachytherapy. In order to describe the absorbed dose in water around a brachy-therapy point source, Meisberger provided a polynomial fit for different isotopes taking into account the effect of attenuation and scattering. The validity of the Meisberger coefficients is restricted to distances up to 10 cm from the source, which is regarded to be satisfactory for most brachytherapy applications. However, for more distant organs it may lead to errors in calculated absorbed dose. For this reason dose measurements have been performed in air and in water around a high activity 60Co source used in high dose rate brachytherapy. Measurements were carried out to distances of 20 cm, using ionization chambers. These data show that at a distance of about 15 cm the amount of scattered radiation virtually equals the amount of primary radiation. This emphasizes the contribution of scattered radiation to the dose in healthy tissue far from the target volume, even with relatively high energy photon radiation of 60Co. It is also shown that the Meisberger data as well as the approach of Van Kleffens and Star [Int. J. Radiat. Oncol. Phys. 5, 557–563 (1979)] lead to significant errors in absorbed dose between distances of 10 and 20 cm from the source. In addition to these measurements, the Monte Carlo code has been used to calculate separately primary dose and scattered dose from a cobalt point source. The calculated results agree with the experimental data within 1% for a most distant dose scoring region. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:549 / 554
页数:6
相关论文
共 29 条
  • [1] Monte Carlo dosimetric study of the Flexisource Co-60 high dose rate source
    Vijande, Javier
    Granero, Domingo
    Perez-Calatayud, Jose
    Ballester, Facundo
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2012, 4 (01) : 34 - 44
  • [2] An analytic approach to the dosimetry of a new BEBIG Co-60 high-dose-rate brachytherapy source
    Bhola, Subhalaxmi
    Selvam, T. Palani
    Sridhar, Sahoo
    Vishwakarma, Ramkrishna S.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (03) : 129 - 137
  • [3] Quality assurance tool for determination of position and transit time of a Co-60 source in high dose rate brachytherapy
    Piriz, Gustavo H.
    Ortega-Spina, Henry G.
    Gonzalez-Sprinberg, Gabriel A.
    APPLIED RADIATION AND ISOTOPES, 2021, 178
  • [4] Monte Carlo dose calculation of GZP6 Co-60 stepping source based on a matrix shift technique
    Toossi, Mohammad Taghi Bahreyni
    Abdollahi, Malihe
    Ghorbani, Mahdi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2011, 16 (01) : 10 - 13
  • [5] Monte Carlo dosimetric characterization of the Flexisource Co-60 high-dose-rate brachytherapy source using PENELOPE
    Almansa, Julio F.
    Guerrero, Rafael
    Torres, Javier
    Lallena, Antonio M.
    BRACHYTHERAPY, 2017, 16 (05) : 1073 - 1080
  • [6] Technical note:: Dosimetric study of a. new Co-60 source used in brachytherapy
    Granero, D.
    Perez-Calatayud, J.
    Ballester, F.
    MEDICAL PHYSICS, 2007, 34 (09) : 3485 - 3488
  • [7] A Monte Carlo investigation of the dose distribution for 60Co high dose rate brachytherapy source in water and in different media
    Badry, Hamza
    Oufni, Lhoucine
    Ouabi, Hmad
    Hirayama, Hideo
    APPLIED RADIATION AND ISOTOPES, 2018, 136 : 104 - 110
  • [8] Dose distribution around a low dose rate 192Ir seed source implanted in the brain
    Ghiassi-Nejad, M
    Jafarizadeh, M
    Rashidi, S
    Pooya, MH
    PHYSICA MEDICA, 2003, 19 (01) : 27 - 30
  • [9] Monte Carlo modeling of Co-60 HDR brachytherapy source in water and in different solid water phantom materials
    Sahoo, S.
    Selvam, T. Palani
    Vishwakarma, R. S.
    Chourasiya, G.
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (01) : 15 - 22
  • [10] Assessment of dose uniformity around high dose rate 192Ir and 60Co stepping sources
    Farhood B.
    Ghorbani M.
    Radiological Physics and Technology, 2017, 10 (4) : 454 - 463