PLANNING OF RADIOTHERAPY FOR PATIENTS WITH HIP-PROSTHESIS

被引:29
作者
ERLANSON, M
FRANZEN, L
HENRIKSSON, R
LITTBRAND, B
LOFROTH, PO
机构
[1] UMEA UNIV,DEPT ONCOL,S-90185 UMEA,SWEDEN
[2] UMEA UNIV,DEPT RADIAT PHYS,S-90185 UMEA,SWEDEN
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1991年 / 20卷 / 05期
关键词
RADIOTHERAPY; HIP PROSTHESIS; TITANIUM DOSE DISTORTIONS;
D O I
10.1016/0360-3016(91)90210-U
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The effects of a hip prosthesis on the dose distribution when treating pelvic cancer have been evaluated. A prosthesis of titanium alloy in a water phantom was used as a model. Photon radiation beams with energies of 6, 20, and 50 MV and with a focus phantom distance of 100 cm were directed against the prosthesis. The dose profiles at different depths were measured with the RFA-7 system. The sphere of the prosthesis was inhomogeneous. The dose in a beam behind the prosthesis was reduced with the order of 10-40% and the largest disortion was behind the shaft. Using an opposed four-field technique with the same weight on all the fields, the dose reduction was as large as 7-12% in a string across the target behind the end of the shaft. With 50% weight on the lateral fields, the dose reduction was 4-8% of the average dose. Such an uneven dose distribution may decrease the curability, and a treatment technique not including the prosthesis should be preferred. Finally, a local increase of the dose close to the prosthesis seems to be of limited practical concern even at the highest energies.
引用
收藏
页码:1093 / 1098
页数:6
相关论文
共 9 条
[1]   DAMAGING EFFECT OF THERAPEUTIC RADIATION ON PROGRAMMABLE PACEMAKERS [J].
ADAMEC, R ;
HAEFLIGER, JM ;
KILLISCH, JP ;
NIEDERER, J ;
JAQUET, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :146-150
[2]   DOSIMETRIC PRECISION REQUIREMENTS IN RADIATION-THERAPY [J].
BRAHME, A .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (05) :379-391
[3]   RADIATION AND DRUG-RESISTANCE OF BREAST-CANCER [J].
FLETCHER, GH .
AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 1984, 7 (06) :617-624
[4]  
HAZUKA MB, 1989, INT J RADIAT ONCOL, V14, P1311
[5]   RADIOTHERAPY TREATMENT PLANNING FOR PATIENTS FITTED WITH PROSTHESES [J].
HUDSON, FR ;
CRAWLEY, MT ;
SAMARASEKERA, M .
BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (679) :603-608
[6]   PACEMAKER FAILURE DUE TO RADIATION-THERAPY [J].
KATZENBERG, CA ;
MARCUS, FI ;
HEUSINKVELD, RS ;
MAMMANA, RB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :156-159
[7]  
RICKNER G, 1987, PHYS MED BIOL, V39, P1109
[8]   AN ANALYSIS OF THE CONTRIBUTION OF COMPUTED-TOMOGRAPHY TO THE TREATMENT OUTCOME IN BLADDER-CANCER [J].
ROTHWELL, RI ;
ASH, DV ;
THOROGOOD, J .
CLINICAL RADIOLOGY, 1985, 36 (04) :369-372
[9]   DOSE PERTURBATION RESULTING FROM GOLD FILLINGS IN PATIENTS WITH HEAD AND NECK CANCERS [J].
THAMBI, V ;
MURTHY, AK ;
ALDER, G ;
KARTHA, PK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (04) :581-582