Postmarket surveillance of 3D-printed implants for sacroiliac joint fusion

被引:13
作者
Cher, Daniel [1 ]
Wroe, Kendrick [2 ]
Reckling, W. Carlton [3 ]
Yerby, Scott [4 ]
机构
[1] SI BONE Inc, Clin Affairs, 471 El Camino Real, Santa Clara, CA 95050 USA
[2] SI BONE Inc, Qual Assurance, Santa Clara, CA USA
[3] SI BONE Inc, Med Affairs, Santa Clara, CA USA
[4] SI BONE Inc, Res & Dev, Santa Clara, CA USA
来源
MEDICAL DEVICES-EVIDENCE AND RESEARCH | 2018年 / 11卷
关键词
permanent implants; quality system; complaint analysis; adverse events; postmarket surveillance; sacroiliac joint fusion;
D O I
10.2147/MDER.S180958
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Government regulations require postmarket surveillance for cleared/approved medical devices. Trend analysis of newly marketed devices may help to confirm device-related safety or uncover other device or procedure-related problems. Methods: Complaints related to the use of 3D-printed triangular titanium implants for sacroiliac joint (SIJ) fusion were compared with those of the prior machined version of the device manufactured with a titanium plasma spray (TPS) coating. Event rates were calculated either by dividing event counts by numbers of surgeries or, for late events, using Kaplan-Meier survival analysis. Results: Three types of complaints with nontrial frequencies were identified. Issues in instruments occurred at a low and constant rate (1.3%). Using Kaplan-Meier analysis, pain-related complaints occurred at a low and similar rate in both groups (<0.5%). The 1-year cumulative probability of surgical revision was low in both the 3D and machined versions of the device (1.5% for machined and 1% for 3D printed, P=0.0408 for difference). No implant breakages or migrations were identified in either group, and overall rates were similar to a previously published report. Conclusion: The 3D-printed version of triangular titanium implant was associated with complaint and adverse event rates similar to those for the prior machined version of the device.
引用
收藏
页码:337 / 343
页数:7
相关论文
共 22 条
[1]  
[Anonymous], 2013, MEDDEV 2 12 1 REV 8
[2]   RECOGNIZING SPECIFIC CHARACTERISTICS OF NONSPECIFIC LOW-BACK-PAIN [J].
BERNARD, TN ;
KIRKALDYWILLIS, WH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1987, (217) :266-280
[3]   mplant survivorship analysis after minimally invasive sacroiliac joint fusion using the iFuse Implant System((R)) [J].
Cher, Daniel J. ;
Reckling, W. Carlton ;
Capobianco, Robyn A. .
MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2015, 8 :485-492
[4]   Predictors of Outcome in Conservative and Minimally Invasive Surgical Management of Pain Originating From the Sacroiliac Joint A Pooled Analysis [J].
Dengler, Julius ;
Duhon, Bradley ;
Whang, Peter ;
Frank, Clay ;
Glaser, John ;
Sturesson, Bengt ;
Garfin, Steven ;
Cher, Daniel ;
Rendahl, Aaron ;
Polly, David .
SPINE, 2017, 42 (21) :1664-1673
[5]  
Dengler J, 2017, PAIN PHYSICIAN, V20, P537
[6]   Revision Surgery Following Operations for Lumbar Stenosis [J].
Deyo, Richard A. ;
Martin, Brook I. ;
Kreuter, William ;
Jarvik, Jeffrey G. ;
Angier, Heather ;
Mirza, Sohail K. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A (21) :1979-1986
[7]   Triangular Titanium Implants for Minimally Invasive Sacroiliac Joint Fusion: 2-Year Follow-Up from a Prospective Multicenter Trial [J].
Duhon, Bradley S. ;
Bitan, Fabien ;
Lockstadt, Harry ;
Kovalsky, Don ;
Cher, Daniel ;
Hillen, Travis .
INTERNATIONAL JOURNAL OF SPINE SURGERY, 2016, 10
[8]  
FDA, MAUDE MAN US FAC DEV
[9]  
Hallab Nadim James, 2004, Dermatitis, V15, P82, DOI 10.2310/6620.2004.03054
[10]  
Hallab NJ, 2001, J BIOMED MATER RES, V56, P427, DOI 10.1002/1097-4636(20010905)56:3<427::AID-JBM1112>3.0.CO