Stability-Indicating High-Performance Liquid Chromatography Assay for the Determination of Sulthiame in Pharmaceutical Dosage Forms

被引:1
|
作者
Haidar, Ammar [1 ]
Kabiche, Sofiane [1 ]
Majoul, Elyes [1 ]
Balde, Issa-Bella [1 ]
Fontan, Jean-Eudes [1 ]
Cisternino, Salvatore [1 ]
Schlatter, Joel [1 ]
机构
[1] Hop Univ Paris Seine St Denis, Hop Jean Verdier, APHP, Serv Pharm, Bondy, France
来源
关键词
sulthiame; liquid chromatography; stability indicating; forced degradation;
D O I
10.4137/ACI.S38656
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A stability-indicating assay by reversed-phase high performance liquid chromatography method was developed and validated for the determination of sulthiame (STM). The chromatographic separation was achieved on a reversed-phase NovaPack C18 column and an isocratic mobile phase consisting of deionized water: methanol (70: 30, v/v). The flow rate was 1.0 mL/min (ultraviolet detection at 210 nm). The STM was separated within 2.83 min. The linearity of the method was demonstrated in the range of 20.0-200.0 mu g/mL and a coefficient of determination of r(2) = 0.9999. The limits of detection and quantification were 4.2 and 9.5 mu g/mL, respectively. The intraday and interday precisions were less than 1%. Accuracy of the method ranged from 98.3% to 101.7%, with a relative standard deviation of. 1%. STM was degraded by accelerated breakdown in alkaline, acidic, or oxidative stress conditions. This method allows accurate and reliable determination of STM for drug stability assay in pharmaceutical studies.