Enhanced resolution pulse-echo imaging with stabilized pulses

被引:6
|
作者
Chen, Shujie [1 ]
Parker, Kevin J. [1 ]
机构
[1] Univ Rochester, Dept Elect & Comp Engn, Hopeman Engn Bldg 203,POB 270126, Rochester, NY 14627 USA
关键词
super-resolution; ultrasound images; lateral resolution enhancement; deconvolution; Z-transform; stable inverse filters;
D O I
10.1117/1.JMI.3.2.027003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Many pulse-echo imaging systems use focused beams to improve lateral resolution. The beam width is determined by the choice of source and apodization function, the frequency, and the physics of focusing. Postprocessing strategies to improve lateral resolution can be limited by the need for conditioning the mathematics of inverse filtering, due to instabilities. We present an analysis that defines key constraints on sampled versions of lateral beampatterns. Within these constraints are useful symmetric beampatterns, which, when properly sampled, can have a stable inverse filter. A framework for analysis and processing is described and applied to phantoms and tissues to demonstrate the improvements that can be realized. (C) 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:13
相关论文