PNEUMATONOGRAPH - LABORATORY STUDY

被引:30
作者
MOSES, RA [1 ]
GRODZKI, WJ [1 ]
机构
[1] WASHINGTON UNIV,SCH MED,OSCAR JOHNSON INST,ST LOUIS,MO 63110
关键词
D O I
10.1001/archopht.1979.01020010291023
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The scale reading of the pneumatonograph (PTG) is dependent on the force of application to the cornea. Since this force is relatively greater at low intraocular pressure, the scale is compressed, yielding an overestimate of intraocular pressure in physiologic range. When the PTG is held vertically, the weight of the plunger is added to the gas propulsion force and the compression of the scale is accentuated. When a 10-g plunger load is added, as is advocated for tonography, the scale compression is further accentuated. We conclude that the PTG scale is compressed, the PTG is position-sensitive, and that the PTG is totally unsuited to tonography. The PTG is not an applanation tonometer and should not be confused with applanation tonometers. © 1979, American Medical Association. All rights reserved.