CLINICAL AND HEMODYNAMIC PERFORMANCE OF THE 19-MM CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS

被引:18
作者
BOJAR, RM
RASTEGAR, H
PAYNE, DD
MACK, CA
SCHWARTZ, SL
机构
[1] TUFTS UNIV NEW ENGLAND MED CTR,DEPT CARDIOL,BOSTON,MA 02111
[2] TUFTS UNIV,SCH MED,BOSTON,MA 02111
关键词
D O I
10.1016/0003-4975(95)90032-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Because of concerns about the hemodynamic performance of 19-mm porcine valves, we retrospectively reviewed the clinical results and echocardiographic studies of 52 consecutive patients who received a 19-mm Carpentier-Edwards porcine bioprosthesis (model 2625) for aortic valve replacement from 1986 through 1991. Nearly 87% of the patients were women, the mean age was 69 years, and the mean body surface area was 1.63 +/- 0.27 m2. Seventy-three percent of the patients had pure aortic stenosis, 96% were in New York Heart Association classes III and IV, and 56% underwent urgent or emergent operation. Overall hospital mortality was 7.7% with a late mortality of 8.3% at a mean follow-up of 25 +/- 18 months. No patient experienced a valve-related complication, and 95% of surviving patients were in New York Heart Association classes I and II. Two-dimensional and Doppler echocardiography performed during the first postoperative week revealed a maximal instantaneous gradient of 44.7 +/- 13.0 mm Hg. In 43 patients for whom additional data were available, the mean gradient was 26.4 +/- 8.2 mm Hg with an effective orifice area of 0.85 +/- 0.18 cm2. This study defines the normal range of postoperative gradients across the 19-mm Carpentier-Edwards porcine valve and demonstrates that patients receiving this valve can achieve significant clinical improvement despite the presence of high transvalvular gradients measured by echocardiography.
引用
收藏
页码:1141 / 1147
页数:7
相关论文
共 27 条
  • [1] ARIS A, 1992, J THORAC CARDIOV SUR, V103, P1074
  • [2] DOPPLER ASSESSMENT OF PROSTHETIC VALVE ORIFICE AREA - AN INVITRO STUDY
    BAUMGARTNER, H
    KHAN, SS
    DEROBERTIS, M
    CZER, LS
    MAURER, G
    [J]. CIRCULATION, 1992, 85 (06) : 2275 - 2283
  • [3] EFFECT OF PROSTHETIC AORTIC-VALVE DESIGN ON THE DOPPLER-CATHETER GRADIENT CORRELATION - AN INVITRO STUDY OF NORMAL ST-JUDE, MEDTRONIC-HALL, STARR-EDWARDS AND HANCOCK VALVES
    BAUMGARTNER, H
    KHAN, S
    DEROBERTIS, M
    CZER, L
    MAURER, G
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (02) : 324 - 332
  • [4] BOJAR RM, 1989, J THORAC CARDIOV SUR, V98, P1087
  • [5] BOVE EL, 1985, J THORAC CARDIOV SUR, V90, P750
  • [6] DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION
    CHAFIZADEH, ER
    ZOGHBI, WA
    [J]. CIRCULATION, 1991, 83 (01) : 213 - 223
  • [7] CHAMBERS J, 1992, J AM COLL CARDIOL, V21, P398
  • [8] EVALUATION OF NORMAL PROSTHETIC VALVE FUNCTION BY DOPPLER ECHOCARDIOGRAPHY
    COOPER, DM
    STEWART, WJ
    SCHIAVONE, WA
    LOMBARDO, HP
    LYTLE, BW
    LOOP, FD
    SALCEDO, EE
    [J]. AMERICAN HEART JOURNAL, 1987, 114 (03) : 576 - 582
  • [9] COSGROVE DM, 1985, CIRCULATION, V72, P146
  • [10] Craver J M, 1979, Circulation, V60, P93