CLINICAL AND HEMODYNAMIC PERFORMANCE OF THE 19-MM CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS

被引:18
作者
BOJAR, RM
RASTEGAR, H
PAYNE, DD
MACK, CA
SCHWARTZ, SL
机构
[1] TUFTS UNIV NEW ENGLAND MED CTR,DEPT CARDIOL,BOSTON,MA 02111
[2] TUFTS UNIV,SCH MED,BOSTON,MA 02111
关键词
D O I
10.1016/0003-4975(95)90032-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Because of concerns about the hemodynamic performance of 19-mm porcine valves, we retrospectively reviewed the clinical results and echocardiographic studies of 52 consecutive patients who received a 19-mm Carpentier-Edwards porcine bioprosthesis (model 2625) for aortic valve replacement from 1986 through 1991. Nearly 87% of the patients were women, the mean age was 69 years, and the mean body surface area was 1.63 +/- 0.27 m2. Seventy-three percent of the patients had pure aortic stenosis, 96% were in New York Heart Association classes III and IV, and 56% underwent urgent or emergent operation. Overall hospital mortality was 7.7% with a late mortality of 8.3% at a mean follow-up of 25 +/- 18 months. No patient experienced a valve-related complication, and 95% of surviving patients were in New York Heart Association classes I and II. Two-dimensional and Doppler echocardiography performed during the first postoperative week revealed a maximal instantaneous gradient of 44.7 +/- 13.0 mm Hg. In 43 patients for whom additional data were available, the mean gradient was 26.4 +/- 8.2 mm Hg with an effective orifice area of 0.85 +/- 0.18 cm2. This study defines the normal range of postoperative gradients across the 19-mm Carpentier-Edwards porcine valve and demonstrates that patients receiving this valve can achieve significant clinical improvement despite the presence of high transvalvular gradients measured by echocardiography.
引用
收藏
页码:1141 / 1147
页数:7
相关论文
共 27 条
[1]  
ARIS A, 1992, J THORAC CARDIOV SUR, V103, P1074
[2]   DOPPLER ASSESSMENT OF PROSTHETIC VALVE ORIFICE AREA - AN INVITRO STUDY [J].
BAUMGARTNER, H ;
KHAN, SS ;
DEROBERTIS, M ;
CZER, LS ;
MAURER, G .
CIRCULATION, 1992, 85 (06) :2275-2283
[3]   EFFECT OF PROSTHETIC AORTIC-VALVE DESIGN ON THE DOPPLER-CATHETER GRADIENT CORRELATION - AN INVITRO STUDY OF NORMAL ST-JUDE, MEDTRONIC-HALL, STARR-EDWARDS AND HANCOCK VALVES [J].
BAUMGARTNER, H ;
KHAN, S ;
DEROBERTIS, M ;
CZER, L ;
MAURER, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (02) :324-332
[4]  
BOJAR RM, 1989, J THORAC CARDIOV SUR, V98, P1087
[5]  
BOVE EL, 1985, J THORAC CARDIOV SUR, V90, P750
[6]   DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION [J].
CHAFIZADEH, ER ;
ZOGHBI, WA .
CIRCULATION, 1991, 83 (01) :213-223
[7]  
CHAMBERS J, 1992, J AM COLL CARDIOL, V21, P398
[8]   EVALUATION OF NORMAL PROSTHETIC VALVE FUNCTION BY DOPPLER ECHOCARDIOGRAPHY [J].
COOPER, DM ;
STEWART, WJ ;
SCHIAVONE, WA ;
LOMBARDO, HP ;
LYTLE, BW ;
LOOP, FD ;
SALCEDO, EE .
AMERICAN HEART JOURNAL, 1987, 114 (03) :576-582
[9]  
COSGROVE DM, 1985, CIRCULATION, V72, P146
[10]  
Craver J M, 1979, Circulation, V60, P93