MEASUREMENT OF PROTEIN USING FLOW-INJECTION ANALYSIS WITH BICINCHONINIC ACID

被引:30
作者
DAVIS, LC
RADKE, GA
机构
[1] Department of Biochemistry, Kansas State University, Manhattan
关键词
bicinchoninic acid; flow injection analysis;
D O I
10.1016/0003-2697(87)90665-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We have used the bicinchoninic acid reagent developed by Pierce Chemical Co. to measure proteins in a simple flow injection analyzer. The sensitivity is comparable to that of the Lowry method and no pipetting of reagents is needed. Results are obtained in less than 1 min and samples may be run at a rate of 60/h. The response is linear over a range of protein concentration (0-10 μg) and sample size (5-20 μl) convenient for most analytical requirements. A peristaltic pump, a controlled-temperature water bath, and a spectrophotometer with flow cuvette are the only special apparatus required. © 1987.
引用
收藏
页码:152 / 156
页数:5
相关论文
共 7 条
[1]   MICRO-DETERMINATION OF NEUTRAL AND AMINO-SUGARS FOUND IN GLYCOPROTEINS [J].
BARR, J ;
NORDIN, P .
ANALYTICAL BIOCHEMISTRY, 1980, 108 (02) :313-319
[3]  
DAVIS LC, 1985, CURRENT TOPICS PLANT, V4, P235
[4]   ON THE CHEMICAL BASIS OF THE LOWRY PROTEIN DETERMINATION [J].
LEGLER, G ;
MULLERPLATZ, CM ;
MENTGESHETTKAMP, M ;
PFLIEGER, G ;
JULICH, E .
ANALYTICAL BIOCHEMISTRY, 1985, 150 (02) :278-287
[5]  
LOWRY OH, 1951, J BIOL CHEM, V193, P265
[6]   LOWRY PROTEIN DETERMINATION BY AUTOMATED FLOW-INJECTION ANALYSIS FOR BOVINE SERUM-ALBUMIN AND HEPATITIS-B SURFACE-ANTIGEN [J].
SALERNO, RA ;
ODELL, C ;
CYANOVICH, N ;
BUBNIS, BP ;
MORGES, W ;
GRAY, A .
ANALYTICAL BIOCHEMISTRY, 1985, 151 (02) :309-314
[7]   MEASUREMENT OF PROTEIN USING BICINCHONINIC ACID [J].
SMITH, PK ;
KROHN, RI ;
HERMANSON, GT ;
MALLIA, AK ;
GARTNER, FH ;
PROVENZANO, MD ;
FUJIMOTO, EK ;
GOEKE, NM ;
OLSON, BJ ;
KLENK, DC .
ANALYTICAL BIOCHEMISTRY, 1985, 150 (01) :76-85