Construction of realistic phantoms from patient images and a commercial three-dimensional printer

被引:28
作者
Leng, Shuai [1 ]
Chen, Baiyu [1 ]
Vrieze, Thomas [1 ]
Kuhlmann, Joel [2 ]
Yu, Lifeng [1 ]
Alexander, Amy [1 ]
Matsumoto, Jane [1 ]
Morris, Jonathan [1 ]
McCollougha, Cynthia H. [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Mayo Clin, Div Engn, 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
phantom; liver; brain; three-dimensional printer; computed tomography; image quality;
D O I
10.1117/1.JMI.3.3.033501
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to use three-dimensional (3-D) printing techniques to construct liver and brain phantoms having realistic pathologies, anatomic structures, and heterogeneous backgrounds. Patient liver and head computed tomography (CT) images were segmented into tissue, vessels, liver lesion, white and gray matter, and cerebrospinal fluid (CSF). Stereolithography files of each object were created and imported into a commercial 3-D printer. Printing materials were assigned to each object after test scans, which showed that the printing materials had CT numbers ranging from 70 to 121 HU at 120 kV. Printed phantoms were scanned on a CT scanner and images were evaluated. CT images of the liver phantom had measured CT numbers of 77.8 and 96.6 HU for the lesion and background, and 137.5 to 428.4 HU for the vessels channels, which were filled with iodine solutions. The difference in CT numbers between lesions and background (18.8 HU) was representative of the low-contrast values needed for optimization tasks. The liver phantom background was evaluated with Haralick features and showed similar texture between patient and phantom images. CT images of the brain phantom had CT numbers of 125, 134, and 108 HU for white matter, gray matter, and CSF, respectively. The CT number differences were similar to those in patient images. (C) 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:7
相关论文
共 32 条
[1]   Contrast-to-Noise Ratio and Low-Contrast Object Resolution on Full- and Low-Dose MDCT: SAFIRE Versus Filtered Back Projection in a Low-Contrast Object Phantom and in the Liver [J].
Baker, Mark E. ;
Dong, Frank ;
Primak, Andrew ;
Obuchowski, Nancy A. ;
Einstein, David ;
Gandhi, Namita ;
Herts, Brian R. ;
Purysko, Andrei ;
Remer, Erick ;
Vachani, Neil .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (01) :8-18
[2]   Application of the noise power spectrum in modern diagnostic MDCT: Part II. Noise power spectra and signal to noise [J].
Boedeker, K. L. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4047-4061
[3]   Evaluating iterative reconstruction performance in computed tomography [J].
Chen, Baiyu ;
Giraldo, Juan Carlos Ramirez ;
Solomon, Justin ;
Samei, Ehsan .
MEDICAL PHYSICS, 2014, 41 (12)
[4]   Effect of Reduced Radiation Exposure and Iterative Reconstruction on Detection of LowContrast Low-A ttenuation Lesions in an Anthropomorphic Liver Phantom: An 18-Reader Study [J].
Goenka, Ajit H. ;
Herts, Brian R. ;
Obuchowski, Nancy A. ;
Primak, Andrew N. ;
Dong, Frank ;
Karim, Wadih ;
Baker, Mark E. .
RADIOLOGY, 2014, 272 (01) :154-163
[5]   Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study [J].
Hara, Amy K. ;
Paden, Robert G. ;
Silva, Alvin C. ;
Kujak, Jennifer L. ;
Lawder, Holly J. ;
Pavlicek, William .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (03) :764-771
[6]   TEXTURAL FEATURES FOR IMAGE CLASSIFICATION [J].
HARALICK, RM ;
SHANMUGAM, K ;
DINSTEIN, I .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1973, SMC3 (06) :610-621
[7]   Comparison between human and model observer performance in low- contrast detection tasks in CT images: application to images reconstructed with filtered back projection and iterative algorithms [J].
Hernandez-Giron, I. ;
Calzado, A. ;
Geleijns, J. ;
Joemai, R. M. S. ;
Veldkamp, W. J. H. .
BRITISH JOURNAL OF RADIOLOGY, 2014, 87 (1039)
[8]   Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements [J].
Kalender, WA ;
Wolf, H ;
Suess, C .
MEDICAL PHYSICS, 1999, 26 (11) :2248-2253
[9]   Computed Tomography Radiation Dose Optimization: Scanning Protocols and Clinical Applications of Automatic Exposure Control [J].
Kalra, Mannudeep K. ;
Naz, Nausheen ;
Rizzo, Stefania M. R. ;
Blake, Michael A. .
CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2005, 34 (05) :171-181
[10]   Penalized-likelihood sinogram restoration for computed tomography [J].
La Riviere, Patrick J. ;
Bian, Junguo ;
Vargas, Phillip A. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (08) :1022-1036