A REAPPRAISAL OF VENTRICULAR-FIBRILLATION THRESHOLD TESTING

被引:19
|
作者
CHA, YM
PETERS, BB
BIRGERSDOTTERGREEN, U
CHEN, PS
机构
[1] UNIV CALIF SAN DIEGO, DEPT MED, DIV CARDIOL, BASIC ARRHYTHMIA LAB, LA JOLLA, CA 92093 USA
[2] VET ADM MED CTR, SAN DIEGO, CA 92161 USA
来源
AMERICAN JOURNAL OF PHYSIOLOGY | 1993年 / 264卷 / 03期
关键词
UP-DOWN ALGORITHM; DOSE-RESPONSE CURVE; VULNERABILITY; ELECTROPHYSIOLOGY;
D O I
10.1152/ajpheart.1993.264.3.H1005
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
To test the hypothesis that ventricular fibrillation (VF) threshold testing is a probability function, 12 open-chest dogs were studied. The VF thresholds were tested by scanning the T wave with either the single premature stimulus method or with the train stimulus method. The dose-response curve method was used to determine the probability of inducing VF with different strengths of premature stimulation. Conventional methods and up-down methods were also used to test the VF threshold. The results showed that the VF threshold is a probability function. The conventional method VF threshold for the single premature stimulus and for the train stimulus methods corresponded to the current strength associated with a 23.4 +/- 13.2 and a 33.2 +/-19.5% probability of VF (P < 0.05), respectively. In comparison, the triplicate VF threshold determined by the up-down algorithm method resulted in values that were not significantly different from a 50% probability of inducing VF, with a good correlation (r = 0.90, P < 0.001 for single and r = 0.89, P = 0.003 for the train stimulus method). We conclude that 1) The VF threshold is a probability function, and 2) the triplicate VF threshold determined by the up-down algorithm method is the best alternative to the dose-response curve method in estimating 50% probability of inducing VF.
引用
收藏
页码:H1005 / H1010
页数:6
相关论文
共 50 条
  • [41] Ventricular response as a predictor of the termination of sustained paroxysmal atrial fibrillation
    Huang, Jen-Hung
    Lin, Yung-Kuo
    Hsieh, Ming-Hsiung
    Chen, Shih-Ann
    Chen, Yi-Jen
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2024, 47 (06): : 843 - 852
  • [42] Idiopathic Ventricular Fibrillation in a 29-Year-Old Man
    Beach, Leila Y.
    Goldschlager, Nora
    Moss, Joshua D.
    Scheinman, Melvin M.
    CIRCULATION, 2017, 136 (01) : 112 - 114
  • [43] Fast Fourier transform analysis of ventricular fibrillation intervals to predict ventricular refractoriness and its spatial dispersion
    Wang, LX
    Li, CY
    Yong, AC
    Kilpatrick, D
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (12): : 2588 - 2595
  • [44] Heterogeneity of Left Ventricular Signal Characteristics in Response to Acute Vagal Stimulation during Ventricular Fibrillation in Dogs
    Nazeri, Alireza
    Elayda, MacArthur A.
    Dragnev, Lubomir
    Frank, Christopher M.
    Qu, Jihong
    Afonso, Valtino X.
    Rasekh, Abdi
    Saeed, Mohammad
    Cheng, Jie
    Shuraih, Mossaab
    Massumi, Ali
    Razavi, Mehdi
    TEXAS HEART INSTITUTE JOURNAL, 2011, 38 (06) : 621 - 626
  • [45] Chemical ablation of the left ventricular endocardium reduces ventricular fibrillation inducibility in acute ischemic canine heart
    Nitta, Takashi
    Asano, Tetsuo
    Yajima, Toshimi
    Otsuka, Toshiaki
    Ishii, Yosuke
    JOURNAL OF ARRHYTHMIA, 2024, 40 (02) : 325 - 332
  • [46] Prevention of Ventricular Fibrillation Episodes in Brugada Syndrome by Catheter Ablation over the Anterior Right Ventricular Epicardium
    Nademanee, Koonlawee
    Veerakul, Gumpanart
    Chandanamattha, Pakorn
    Chaothawee, Lertlak
    Ariyachaipanich, Aekarach
    Jirasirirojanakorn, Kriengkrai
    Likittanasombat, Khanchit
    Bhuripanyo, Kiertijai
    Ngarmukos, Tachapong
    CIRCULATION, 2010, 122 (21)
  • [47] Significant Reduction of Atrial Defibrillation Threshold and Inducibility by Catheter Ablation of Atrial Fibrillation
    Hwang, Eui-Seock
    Nam, Gi Byung
    Joung, Boyoung
    Park, Junbeom
    Lee, Jihei Sara
    Shim, Jaemin
    Uhm, Jae-Sun
    Lee, Moon Hyoung
    Pak, Hui-Nam
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (12): : 1428 - 1435
  • [48] Regression of left ventricular hypertrophy with captopril restores normal ventricular action potential duration, dispersion of refractoriness, and vulnerability to inducible ventricular fibrillation
    Rials, SJ
    Wu, Y
    Xu, XP
    Filart, RA
    Marinchak, RA
    Kowey, PR
    CIRCULATION, 1997, 96 (04) : 1330 - 1336
  • [49] Systematic Review of Defibrillation Threshold Testing at De Novo Implantation
    Phan, Kevin
    Ha, Hakeem
    Kabunga, Peter
    Kilborn, Michael J.
    Toal, Edward
    Sy, Raymond W.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2016, 9 (04)
  • [50] Dosimetry for Ventricular Fibrillation Risk with Short Electrical Pulses: History and Future
    Kroll, Mark W.
    Panescu, Dorin
    Hirtler, Reinhard
    Koch, Michael
    Andrews, Christopher J.
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 1788 - 1794