ACCURACY OF DEXA MEASUREMENT OF BONE-MINERAL DENSITY AFTER TOTAL HIP-ARTHROPLASTY

被引:124
作者
COHEN, B
RUSHTON, N
机构
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1995年 / 77B卷 / 03期
关键词
D O I
10.1302/0301-620X.77B3.7744941
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Dual-energy X-ray absorptiometry (DEXA) is increasingly used to measure changes in bone mineral density (BMD) around femoral prostheses after total hip arthroplasty, We have studied the factors which affect the accuracy of these measurements. The coefficient of variation was <2% using a hydroxyapatite phantom, 2.7% in an anthropomorphic phantom specimen, and <1% in repeated measurements on implanted cadaver femora, The precision did not vary with different implant materials or designs. In patients we found a mean precision error of 2.7% to 3.4%, The most significant factor affecting reproducibility was rotation of the femur. We conclude that DEXA is a precise method of measurement for small changes in BMD around femoral implants, but that correct and careful positioning of patients is essential to obtain reliable results.
引用
收藏
页码:479 / 483
页数:5
相关论文
共 18 条
  • [1] BOBYN JD, 1992, CLIN ORTHOP RELAT R, P79
  • [2] BRAND RA, 1985, CLIN ORTHOP RELAT R, V192, P237
  • [3] RELATIONSHIPS BETWEEN LOADING HISTORY AND FEMORAL CANCELLOUS BONE ARCHITECTURE
    CARTER, DR
    ORR, TE
    FYHRIE, DP
    [J]. JOURNAL OF BIOMECHANICS, 1989, 22 (03) : 231 - 244
  • [4] COHEN B, 1995, IN PRESS J BONE B S1, V77
  • [5] X-RAY DUAL-PHOTON ABSORPTIOMETRY - A NEW METHOD FOR THE MEASUREMENT OF BONE-DENSITY
    CULLUM, ID
    ELL, PJ
    RYDER, JP
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (739) : 587 - 592
  • [6] A QUANTITATIVE-EVALUATION OF PERIPROSTHETIC BONE-REMODELING AFTER CEMENTLESS TOTAL HIP-ARTHROPLASTY
    ENGH, CA
    MCGOVERN, TF
    BOBYN, JD
    HARRIS, WH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) : 1009 - 1020
  • [7] POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS
    ENGH, CA
    BOBYN, JD
    GLASSMAN, AH
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01): : 45 - 55
  • [8] FENNER JAK, 1992, CURRENT RES OSTEOPOR, V2, P35
  • [9] Gruen T., 2008, QUANTITATIVE CHARACT, V207, DOI 10.1520/stp25232s
  • [10] DUAL-ENERGY X-RAY ABSORPTIOMETRY MEASUREMENT OF BONE-MINERAL DENSITY AROUND POROUS-COATED CEMENTLESS FEMORAL IMPLANTS - METHODS AND PRELIMINARY-RESULTS
    KILGUS, DJ
    SHIMAOKA, EE
    TIPTON, JS
    EBERLE, RW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02): : 279 - 287