FUZZY CONTROL CONCEPT FOR A TOTAL ARTIFICIAL-HEART

被引:12
作者
KAUFMANN, R
BECKER, K
NIX, C
REUL, H
RAU, G
机构
[1] Helmholtz-Institute for Biomedical Engineering, RWTH, Aachen
关键词
TAH CONTROL; FUZZY CONTROL; CONTROL SENSORS; IN VITRO TESTING;
D O I
10.1111/j.1525-1594.1995.tb02340.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The development of an electromechanically driven total artificial heart (Helmholtz-TAH) was initiated in 1990. Anatomical fitting, biocompatibility, and automatic physiologic adaptation of pump output are the basic requirements that characterize the overall TAH concept. For evaluation of these features, a TAH labtype was developed. It provides most features of the conceptual artificial heart and supports in vitro testing of energy conversion, pump behavior, structural parts, sensors, and control concepts. A fuzzy controller has been implemented for adaptation of the pump rate to body perfusion demand by left pump chamber filling detection. This controller will be an important element of a future extensive TAH control system. The implementation is supported by a professional fuzzy control development tool that allows on-line and real time optimization of control strategies for dynamic processes. The first experiments proved the feasibility and the advantages of this fuzzy control concept. The first in vitro test results are presented.
引用
收藏
页码:355 / 361
页数:7
相关论文
共 14 条
[1]  
BECKER K, 1994, IN PRESS IEEE ENG ME
[2]   THE HELMHOLTZ TOTAL ARTIFICIAL-HEART LABTYPE [J].
KAUFMANN, R ;
REUL, H ;
RAU, G .
ARTIFICIAL ORGANS, 1994, 18 (07) :537-542
[3]  
KAUFMANN R, 1992, INT J ARTIF ORGANS, V8, P481
[4]  
KNIERBEIN B, 1992, INT J ARTIF ORGANS, V1, P40
[5]  
KRUSE R., 1993, FUZZY SYSTEME
[6]  
KUNG RTV, 1991, ARTIFICIAL HEART, V3, P173
[7]   EXPERIMENT IN LINGUISTIC SYNTHESIS WITH A FUZZY LOGIC CONTROLLER [J].
MAMDANI, EH ;
ASSILIAN, S .
INTERNATIONAL JOURNAL OF MAN-MACHINE STUDIES, 1975, 7 (01) :1-13
[8]  
MINAMI K, 1993, HEART REPLACEMENT, P379
[9]  
PANTALOS GM, 1993, ARTIF ORGANS, V10, P826
[10]  
RAU G, 1994, ARTIF ORGANS, V19, P105