Quantitative performance characterization of image quality and radiation dose for a CS 9300 dental cone beam computed tomography machine

被引:10
作者
Abouei, Elham [1 ]
Lee, Sierra [2 ]
Ford, Nancy L. [1 ,2 ]
机构
[1] Univ British Columbia, Dept Phys & Astron, 6224 Agr Rd, Vancouver, BC V6T 1Z1, Canada
[2] Univ British Columbia, Dept Oral Biol & Med Sci, 2199 Webrook Mall, Vancouver, BC V6T 1Z3, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
image quality; image evaluation; computer tomography; imaging;
D O I
10.1117/1.JMI.2.4.044002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper aims to characterize the radiation dose and image quality (IQ) performance of a dental cone beam computed tomography (CBCT) unit over a range of fields of view (FOV). IQ and dose were measured using a Carestream 9300 dental CBCT. Phantoms were positioned in the FOV to imitate clinical positioning. IQ was assessed by scanning a SEDENTEXCT IQ phantom, and images were analyzed in ImageJ. Dose index 1 was obtained using a thimble ionization chamber and SEDENTEXCT DI phantom. Mean gray values agreed within 93.5% to 99.7% across the images, with pixel-to-pixel fluctuations of 6% to 12.5%, with poorer uniformity and increased noise for child protocols. CNR was fairly constant across FOVs, with higher CNR for larger patient settings. The measured limiting spatial resolution agreed well with 10% MTF and bar pattern measurements. Dose was reduced for smaller patient settings within a given FOV; however, smaller FOVs obtained with different acquisition settings did not necessarily result in reduced dose. The use of patient-specific acquisition settings decreased the radiation dose for smaller patients, with minimal impact on the IQ. The full set of IQ and dose measurements is reported to allow dental professionals to compare the different FOV settings for clinical use. (C) The Authors.
引用
收藏
页数:12
相关论文
共 24 条
[1]  
[Anonymous], 2012, CON BEAM CT DENT MAX
[2]   Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom [J].
Bamba, J. ;
Araki, K. ;
Endo, A. ;
Okano, T. .
DENTOMAXILLOFACIAL RADIOLOGY, 2013, 42 (08)
[3]   Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today? [J].
Bruellmann, D. ;
Schulze, R. K. W. .
DENTOMAXILLOFACIAL RADIOLOGY, 2015, 44 (01)
[4]   Study of the scan uniformity from an i-CAT cone beam computed tomography dental imaging system [J].
Bryant, J. A. ;
Drage, N. A. ;
Richmond, S. .
DENTOMAXILLOFACIAL RADIOLOGY, 2008, 37 (07) :365-374
[5]   Measuring absorbed dose for i-CAT CBCT examinations in child, adolescent and adult phantoms [J].
Choi, E. ;
Ford, N. L. .
DENTOMAXILLOFACIAL RADIOLOGY, 2015, 44 (06)
[6]   Relationship between physical factors and subjective image quality of cone-beam computed tomography images according to diagnostic task [J].
Choi, Jin-Woo ;
Lee, Sam-Sun ;
Choi, Soon-Chul ;
Heo, Min-Suk ;
Huh, Kyung-Heo ;
Yi, Won-Jin ;
Kang, Se-Ryong ;
Han, Dong-Hun ;
Kim, Eun-Kyung .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2015, 119 (03) :357-365
[7]   Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: A systematic review of the literature [J].
De Vos, W. ;
Casselman, J. ;
Swennen, G. R. J. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 38 (06) :609-625
[8]   MSCT versus CBCT: evaluation of high-resolution acquisition modes for dento-maxillary and skull-base imaging [J].
Dillenseger, Jean-Philippe ;
Matern, Jean-Francois ;
Gros, Catherine-Isabelle ;
Bornert, Fabien ;
Goetz, Christian ;
Le Minor, Jean-Marie ;
Constantinesco, Andre ;
Choquet, Philippe .
EUROPEAN RADIOLOGY, 2015, 25 (02) :505-515
[9]   A quality assurance phantom for the performance evaluation of volumetric micro-CT systems [J].
Du, Louise Y. ;
Umoh, Joseph ;
Nikolov, Hristo N. ;
Pollmann, Steven I. ;
Lee, Ting-Yim ;
Holdsworth, David W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (23) :7087-7108
[10]  
Gayou O, 2012, J APPL CLIN MED PHYS, V13, P14, DOI 10.1120/jacmp.v13i1.3638