PULSED OSCILLATION TECHNIQUE FOR ASSESSING THE MECHANICAL STATE OF THE DENTAL IMPLANT-BONE INTERFACE

被引:23
作者
KANEKO, T
机构
[1] Research Center, Nikon Corporation, Shinagawa-ku, Tokyo, 140
关键词
DENTAL IMPLANT; INTERFACE; VIBRATION;
D O I
10.1016/0142-9612(91)90050-K
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The sensitivity of a vibrational test for in vivo assessing the interfacial rigidity of a biocompatible dental root implant was studied using models. Aluminium alloy rods of 4 mm diameter and 7 mm long were root implant models. Cyanoacrylate and epoxy adhesives were hard interface models. Soft interfaces were modelled in a soft silicone adhesive, a gel, adhesive tapes and direct loose contact. Blocks of bovine and canine jaw-bones chemically treated and dried were models of a human alveolar bone. It was found that the minimum average thickness of a soft interface layer distinguishable from a hard interface depends on load directions and positions and ranges from 0 to 160-mu-m. The sensitivity was rather low in the direction for which a normal load was applied to the bone, reflecting a mechanical difference of the surrounding bone and/or the interface. Therefore, it is desirable that the assessment by the vibrational test is done in the direction, too, for which a shearing load is applied to the interface.
引用
收藏
页码:555 / 560
页数:6
相关论文
共 23 条
[1]  
DHOEDT B, 1985, DEUT ZAHNAERZTL Z, V40, P113
[2]  
FUJINO M, 1988, 3RD INT C IMPL BIOM, pB33
[3]  
IIJIMA T, 1988, 3RD INT C IMPL BIOM, pB3
[4]  
KAMEGAI T, 1988, 3RD INT C IMPLANTOLO, pB29
[5]   ASSESSMENT OF THE INTERFACIAL RIGIDITY OF BONE IMPLANTS FROM VIBRATIONAL SIGNALS [J].
KANEKO, T .
JOURNAL OF MATERIALS SCIENCE, 1987, 22 (10) :3495-3502
[6]   COMPARISON BETWEEN ACOUSTIC AND MECHANICAL TAPPING METHODS FOR ASSESSING THE INTERFACIAL STATES OF BONE IMPLANTS [J].
KANEKO, T .
JOURNAL OF MATERIALS SCIENCE, 1989, 24 (08) :2820-2824
[7]  
KOHNO S, 1987, QUINTESSENCE, V6, P187
[8]  
Konig M, 1981, Dtsch Zahnarztl Z, V36, P451
[9]  
KUDO K, 1986, JPN J ORAL MAXILLOFA, V32, P2083
[10]  
KUDO K, 1987, BIOCERAMICS DEV CLIN, P193