10 KHZ MICROSECOND PULSED X-RAY GENERATOR UTILIZING A HOT-CATHODE TRIODE WITH VARIABLE DURATIONS FOR BIOMEDICAL RADIOGRAPHY

被引:29
作者
SATO, E
SAGAE, M
TAKAHASHI, K
SHIKODA, A
OIZUMI, T
HAYASI, Y
TAMAKAWA, Y
YANAGISAWA, T
机构
[1] IWATE MED UNIV,MED COMP RES CTR,MORIOKA,IWATE 020,JAPAN
[2] HACHINOHE INST TECHNOL,HACHINOHE 03911,JAPAN
[3] IWATE MED UNIV HOSP,CTR RADIOL SCI,MORIOKA,IWATE 020,JAPAN
[4] IWATE MED UNIV,DEPT RADIOL,MORIOKA,IWATE 020,JAPAN
关键词
HOT-CATHODE TRIODE; HIGH REPETITION RATE; KHZ CINERADIOGRAPHY; MICROSECOND VARIABLE DURATION; PULSED X-RAYS;
D O I
10.1007/BF02512525
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A 10 kHz pulsed X-ray generator utilising a hot-cathode triode in conjunction with a new type of grid control device for controlling X-ray duration is described. The energy-storage condenser was charged up to 70 kV by a power supply, and the electric charges in the condenser were discharged to the X-ray tube repetitively by the grid control device. The maximum values of the grid voltage (negative value), the tube voltage, and the tube current were -1.5 kV, 70 kV, and 0.4 A, respectively. The duration of the flash X-ray pulse was primarily determined by the time constant of the grid control device and the cut-off voltage of thermoelectrons. The X-ray duration was controlled within a region of less than 1 ms; the X-ray intensity with a pulse width of 0.27 ms, a charged voltage of 70 kV, and a peak tube current of 0.4 A was 0.92 mu C kg(-1) at 0.5 m per pulse. The maximum repetition rate was about 10 kHz, and the size of the focal spot was about 3.5 X 3.5 mm.
引用
收藏
页码:295 / 301
页数:7
相关论文
共 16 条
  • [1] DAVANLOO F, 1987, REV SCI INSTRUM, V59, P2103
  • [2] X-RAY FLASH TECHNIQUES
    GERMER, R
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1979, 12 (05): : 336 - 350
  • [3] FLASH X-RAY TUBE FOR DIFFRACTION STUDIES ON A 2-STAGE LIGHT-GAS GUN
    JOHNSON, Q
    MITCHELL, AC
    SMITH, ID
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1980, 51 (06) : 741 - 749
  • [4] SOME CHARACTERISTICS OF A 600 KV FLASH X-RAY TUBE
    MATTSSON, A
    [J]. PHYSICA SCRIPTA, 1972, 5 (1-2): : 99 - &
  • [5] FLASH X-RAY MICROSCOPY
    MCCORKLE, R
    ANGILELLO, J
    COLEMAN, G
    FEDER, R
    LAPLACA, SJ
    [J]. SCIENCE, 1979, 205 (4404) : 401 - 402
  • [6] MEASUREMENT OF FOCAL SPOT SIZE OF DIAGNOSTIC X-RAY TUBES - COMPARISON OF PINHOLE AND RESOLUTION METHODS
    ROBINSON, A
    GRIMSHAW, GM
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1975, 48 (571) : 572 - 580
  • [7] LASER-DRIVEN 3-KA X-RAY SOURCE
    ROTH, S
    FREUND, I
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1978, 49 (01) : 104 - 110
  • [8] KILOHERTZ-RANGE FLASH X-RAY GENERATOR UTILIZING A TRIODE IN CONJUNCTION WITH AN EXTREMELY HOT CATHODE
    SATO, E
    SHIKODA, A
    KIMURA, S
    SAGAE, M
    ISOBE, H
    TAKAHASHI, K
    TAMAKAWA, Y
    YANAGISAWA, T
    HONDA, K
    YOKOTA, Y
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (09) : 2115 - 2120
  • [9] SATO E, 1986, REV SCI INSTRUM, V57, P1399, DOI 10.1063/1.1138609
  • [10] REPETITIVE FLASH X-RAY GENERATOR UTILIZING A SIMPLE DIODE WITH A NEW TYPE OF ENERGY-SELECTIVE FUNCTION
    SATO, E
    KIMURA, S
    KAWASAKI, S
    ISOBE, H
    TAKAHASHI, K
    TAMAKAWA, Y
    YANAGISAWA, T
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (09) : 2343 - 2348