QUANTIFICATION OF REGURGITANT FLOW-THROUGH BILEAFLET HEART-VALVE PROSTHESES - THEORETICAL AND IN-VITRO STUDIES

被引:2
作者
CAPE, EG
NANDA, NC
YOGANATHAN, AP
机构
[1] UNIV ALABAMA,SCH MED,DIV CARDIOVASC DIS,BIRMINGHAM,AL 35233
[2] GEORGIA INST TECHNOL,SCH CHEM ENGN,CARDIOVASC FLUID MECH LAB,ATLANTA,GA 30332
关键词
HEART VALVES; REGURGITATION; DOPPLER ECHOCARDIOGRAPHY; JETS; ULTRASOUND; TURBULENCE; PROSTHESES; ARTIFICIAL HEART VALVES;
D O I
10.1016/0301-5629(93)90122-5
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A theoretical treatment using turbulent jet theory has yielded a new equation for predicting regurgitant flow through bileaflet heart valve prostheses, the most commonly implanted mechanical valve design. Previously reported techniques assuming an axisymmetric jet are not applicable to the slot-like orifices presented in these valves. The equations were therefore rederived in the context of the prosthetic valve geometry. The purpose of this study was to develop such a method and demonstrate its applicability in principle by using in vitro models. The method was validated under both steady and pulsatile flow conditions. Having derived a method geometrically specific to the orifices presented in bileaflet mechanical heart valves, it should be applicable from patient to patient due to the rigid nature of the valve. These idealized in vitro studies, along with the accompanying theoretical derivation, will guide implementation in the clinical setting.
引用
收藏
页码:461 / 468
页数:8
相关论文
共 15 条
[1]  
BLEVINS RD, 1984, APPLIED FLUID DYNAMI, P229
[2]   ADJACENT SOLID BOUNDARIES ALTER THE SIZE OF REGURGITANT JETS ON DOPPLER COLOR FLOW MAPS [J].
CAPE, EG ;
YOGANATHAN, AP ;
WEYMAN, AE ;
LEVINE, RA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (05) :1094-1102
[3]   INCREASED HEART-RATE CAN CAUSE UNDERESTIMATION OF REGURGITANT JET SIZE BY DOPPLER COLOR FLOW MAPPING [J].
CAPE, EG ;
YOGANATHAN, AP ;
LEVINE, RA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (04) :1029-1037
[4]   A NEW METHOD FOR NONINVASIVE QUANTIFICATION OF VALVULAR REGURGITATION BASED ON CONSERVATION OF MOMENTUM - INVITRO VALIDATION [J].
CAPE, EG ;
SKOUFIS, EG ;
WEYMAN, AE ;
YOGANATHAN, AP ;
LEVINE, RA .
CIRCULATION, 1989, 79 (06) :1343-1353
[5]  
CAPE EG, 1990, J BIOMECH, V230, P27
[6]   INVESTIGATIONS OF LAMINAR-FLOW IN A PIPE WITH SUDDEN CONTRACTION OF CROSS-SECTIONAL AREA [J].
DURST, F ;
LOY, T .
COMPUTERS & FLUIDS, 1985, 13 (01) :15-36
[7]   COLOR DOPPLER ASSESSMENT OF HIGH FLOW VELOCITIES USING A NEW TECHNOLOGY - INVITRO AND CLINICAL-STUDIES [J].
FAN, P ;
NANDA, NC ;
COOPER, JW ;
CAPE, E ;
YOGANATHAN, A .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 1990, 7 (06) :763-769
[8]   PATTERNS OF NORMAL TRANSVALVULAR REGURGITATION IN MECHANICAL VALVE PROSTHESES [J].
FLACHSKAMPF, FA ;
OSHEA, JP ;
GRIFFIN, BP ;
GUERRERO, L ;
WEYMAN, AE ;
THOMAS, JD .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (06) :1493-1498
[9]   COLOR DOPPLER ASSESSMENT OF MITRAL REGURGITATION WITH ORTHOGONAL PLANES [J].
HELMCKE, F ;
NANDA, NC ;
HSIUNG, MC ;
SOTO, B ;
ADEY, CK ;
GOYAL, RG ;
GATEWOOD, RP .
CIRCULATION, 1987, 75 (01) :175-183
[10]  
NANDA NC, 1985, DOPPLER ECHOCARDIOGR, P453