FERROELECTRIC TEMPERATURE SENSORS FOR THERMOMETRIC TITRATIONS AND ENTHALPIMETRIC ANALYSIS

被引:4
作者
LYNCH, JA [1 ]
JORDAN, J [1 ]
机构
[1] PENN STATE UNIV,DEPT CHEM,UNIV PK,PA 16802
基金
美国国家科学基金会;
关键词
ENTHALPIMETRY; THERMOMETRIC ANALYSIS; TITRIMETRY; BARIUM TITANATE; FERROELECTRIC TEMPERATURE SENSORS; TEMPERATURE SENSORS;
D O I
10.1016/0003-2670(91)87116-O
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The capabilities of commercially available barium titanate ferroelectric ceramics with switching points of 25 and 0-degrees-C were tested for use in thermometric titrations. Temperature (T) versus resistance (R) data taken from 24 to 26-degrees-C for seven ferroelectrics and from -1 to +1-degrees-C for three others were fitted to an exponential equation of the form R = Aexp(CT), where A is a scaling factor and C is the temperature coefficient of resistance. A ferroelectric sensor (FES) was wired in a Wheatstone bridge in place of the thermistor customarily used in conventional instrumentation. Calculations substantiated that the bridge imbalance potential response approximated linearity within any 1-degrees-C interval, provided that the bridge was balanced in the middle of the span. The response of the FES bridge showed a fourfold enhancement for a given temperature change compared with the thermistor bridge and a concomitant reduction in baseline noise. Analytical capabilities are illustrated by thermometric redox titrations in the concentration range between 3 x 10(-5) and 2 x 10(-3) M.
引用
收藏
页码:59 / 67
页数:9
相关论文
共 27 条