Preclinical evaluation of acoustic radiation force impulse measurements in regions of heterogeneous elasticity

被引:10
作者
Hollerieth, Katharina [1 ]
Gassmann, Bernhard [2 ]
Wagenpfeil, Stefan [3 ]
Moog, Philipp [1 ]
Minh-Truc Vo-Cong [1 ]
Heemann, Uwe [1 ]
Stock, Konrad Friedrich [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Dept Nephrol, Str 22, D-81675 Munich, Germany
[2] Meso Int GmbH, Berlin, Germany
[3] Univ Saarland, Inst Med Biometry Epidemiol & Med Informat, Campus Homburg Saar, Homburg, Germany
关键词
Elasticity imaging techniques; Ultrasonography; Phantoms; imaging; Transducers; Heterogeneity;
D O I
10.14366/usg.16024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to compare the reliability of ultrasound-based shear wave elastography in regions of homogeneous versus heterogeneous elasticity by using two different probes. Methods: Using acoustic radiation force impulse (ARFI) elastography, we measured the shear wave velocity (SWV) in different lesions of an elastography phantom with the convex 4C1 probe and the linear 9L4 probe. The region of interest (ROI) was positioned in such a way that it was partly filled by one of the lesions (0%, 25%, 50%, 75%, and 100%) and partly by the background of the phantom (100%, 75%, 50%, 25%, and 0%, respectively). Results: The success rate was 98.5%. The measured value and the reference value of SWV correlated significantly (r=0.89, P<0.001). Further, a comparison of the two probes revealed that there was no statistical difference in either the mean or the variance values. However, the deviation of SWV from the reference was higher in the case of the 9L4 probe than in the case of the 4C1 probe, both overall and in measurements in which the ROI contained structures of different elasticity (P=0.021 and P=0.002). Taking into account all data, for both probes, we found that there was a greater spread and deviation of the SWV from the reference value when the ROI was positioned in structures having different elastic properties (standard deviation, 0.02 +/- 0.01 m/sec vs. 0.04 +/- 0.04 m/sec; P=0.010; deviation from the reference value, 0.21 +/- 0.12 m/sec vs. 0.38 +/- 0.27 m/sec; P=0.050). Conclusion: Quantitative ARFI elastography was achievable in structures of different elasticity; however, the validity and the reliability of the SWV measurements decreased in comparison to those of the measurements performed in structures of homogeneous elasticity. Therefore, a convex probe is preferred for examining heterogeneous structures.
引用
收藏
页码:345 / 352
页数:8
相关论文
共 29 条
  • [1] Ultrasound Elastography of Lateral Epicondylosis: Clinical Feasibility of Quantitative Elastographic Measurements
    Ahn, Kyung-Sik
    Kang, Chang Ho
    Hong, Suk-Joo
    Jeong, Woong-Kyo
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (05) : 1094 - 1099
  • [2] WFUMB GUIDELINES AND RECOMMENDATIONS FOR CLINICAL USE OF ULTRASOUND ELASTOGRAPHY: PART 2: BREAST
    Barr, Richard G.
    Nakashima, Kazutaka
    Amy, Dominique
    Cosgrove, David
    Farrokh, Andre
    Schafer, Fritz
    Bamber, Jeffrey C.
    Castera, Laurent
    Choi, Byung Ihn
    Chou, Yi-Hong
    Dietrich, Christoph F.
    Ding, Hong
    Ferraioli, Giovanna
    Filice, Carlo
    Friedrich-Rust, Mireen
    Hall, Timothy J.
    Nightingale, Kathryn R.
    Palmeri, Mark L.
    Shiina, Tsuyoshi
    Suzuki, Shinichi
    Sporea, Ioan
    Wilson, Stephanie
    Kudo, Masatoshi
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (05) : 1148 - 1160
  • [3] Non-invasive assessment of liver graft fibrosis by transient elastography after liver transplantation
    Barrault, Camille
    Roudot-Thoraval, Francoise
    Van Nhieu, Jeanne Tran
    Atanasiu, Calina
    Kluger, Michael D.
    Medkour, Fatiha
    Douvin, Catherine
    Mallat, Ariane
    Zafrani, Elie-Serge
    Cherqui, Daniel
    Duvoux, Christophe
    [J]. CLINICS AND RESEARCH IN HEPATOLOGY AND GASTROENTEROLOGY, 2013, 37 (04) : 347 - 352
  • [4] EFSUMB Guidelines and Recommendations on the Clinical Use of Ultrasound Elastography. Part 2: Clinical Applications
    Cosgrove, D.
    Piscaglia, F.
    Bamber, J.
    Bojunga, J.
    Correas, J. -M.
    Gilja, O. H.
    Klauser, A. S.
    Sporea, I.
    Calliada, F.
    Cantisani, V.
    D'Onofrio, M.
    Drakonaki, E. E.
    Fink, M.
    Friedrich-Rust, M.
    Fromageau, J.
    Havre, R. F.
    Jenssen, C.
    Ohlinger, R.
    Saftoiu, A.
    Schaefer, F.
    Dietrich, C. F.
    [J]. ULTRASCHALL IN DER MEDIZIN, 2013, 34 (03): : 238 - 253
  • [5] Acoustic radiation force impulse imaging of the testes: A preliminary experience
    D'Anastasi, M.
    Schneevoigt, B. S.
    Trottmann, M.
    Crispin, A.
    Stief, C.
    Reiser, M. F.
    Clevert, D. A.
    [J]. CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2011, 49 (1-4) : 105 - 114
  • [6] Tissue Quantification With Acoustic Radiation Force Impulse Imaging: Measurement Repeatability and Normal Values in the Healthy Liver
    D'Onofrio, Mirko
    Gallotti, Anna
    Mucelli, Roberto Pozzi
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) : 132 - 136
  • [7] Lower-Limb Vascular Imaging with Acoustic Radiation Force Elastography: Demonstration of In Vivo Feasibility
    Dumont, Douglas
    Dahl, Jeremy
    Miller, Elizabeth
    Allen, Jason
    Fahey, Brian
    Trahey, Gregg
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (05) : 931 - 944
  • [8] WFUMB GUIDELINES AND RECOMMENDATIONS FOR CLINICAL USE OF ULTRASOUND ELASTOGRAPHY: PART 3: LIVER
    Ferraioli, Giovanna
    Filice, Carlo
    Castera, Laurent
    Choi, Byung Ihn
    Sporea, Ioan
    Wilson, Stephanie R.
    Cosgrove, David
    Dietrich, Christoph F.
    Amy, Dominique
    Bamber, Jeffrey C.
    Barr, Richard
    Chou, Yi-Hong
    Ding, Hong
    Farrokh, Andre
    Friedrich-Rust, Mireen
    Hall, Timothy J.
    Nakashima, Kazutaka
    Nightingale, Kathryn R.
    Palmeri, Mark L.
    Schafer, Fritz
    Shiina, Tsuyoshi
    Suzuki, Shinichi
    Kudo, Masatoshi
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (05) : 1161 - 1179
  • [9] Liver Fibrosis in Viral Hepatitis: Noninvasive Assessment with Acoustic Radiation Force Impulse Imaging versus Transient Elastography
    Friedrich-Rust, Mireen
    Wunder, Katrin
    Kriener, Susanne
    Sotoudeh, Fariba
    Richter, Swantje
    Bojunga, Joerg
    Herrmann, Eva
    Poynard, Thierry
    Dietrich, Christoph F.
    Vermehren, Johannes
    Zeuzem, Stefan
    Sarrazin, Christoph
    [J]. RADIOLOGY, 2009, 252 (02) : 595 - 604
  • [10] VISCOELASTIC AND ANISOTROPIC MECHANICAL PROPERTIES OF IN VIVO MUSCLE TISSUE ASSESSED BY SUPERSONIC SHEAR IMAGING
    Gennisson, Jean-Luc
    Deffieux, Thomas
    Mace, Emilie
    Montaldo, Gabriel
    Fink, Mathias
    Tanter, Mickael
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2010, 36 (05) : 789 - 801