FLOW INJECTION-FLAME ATOMIC-ABSORPTION SPECTROMETRIC DETERMINATION OF TRACE LEVELS OF CADMIUM WITH ONLINE PRECONCENTRATION BY MEANS OF COPRECIPITATION

被引:0
作者
MIN, RW [1 ]
HANSEN, EH [1 ]
机构
[1] TECH UNIV DENMARK,DEPT CHEM A,DK-2800 LYNGBY,DENMARK
关键词
TRACE METAL ANALYSIS; ONLINE PRECONCENTRATION BY COPRECIPITATION IN KNOTTED REACTOR; ELUTION BY ORGANIC SOLVENT; FLAME AAS; FLOW INJECTION; ASSAY OF CADMIUM;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Flow injection systems for determination of trace metal ions are described, which comprize on-line preconcentration by coprecipitation with ensuing dissolution and detection by flame atomic absorption spectrometry. Exemplified for the assay of cadmium, which is coprecipitated with iron(II) by means of hexamethylene ammonium hexamethylene dithiocarbamate in an incorporated knotted reactor and subsequently eluted by an organic solvent of methyl isobutyl ketone, two optional FI-approaches for accomplishing this feat are compared and discussed in respect to performance. While one, which is based on a previously published procedure, employs continuous pumping of the eluting agent, the approach suggested herein relies on dissolution of the collected precipitate within a small, discrete organic zone. Because the concentration of analyte thus is increased in the eluent, this concept allows an increased sensitivity, a lower limit of detection (0.074 mu g l(-1); 3 sigma), and an extended dynamic measuring range (0.5-20 mu g l(-1)), albeit implying an increased sample consumption per assay and hence a decreased sampling frequency.
引用
收藏
页码:361 / 371
页数:11
相关论文
共 50 条
[41]   Flow injection on-line displacement sorption preconcentration coupled with flame atomic absorption spectrometry for the determination of trace lead in biological samples [J].
Han De-Man ;
Jia Wen-Ping ;
Liang Hua-Ding ;
Yan Xiu-Ping .
CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2007, 35 (11) :1643-1646
[42]   Efficient lead preconcentration using two chemically functionalized carbon nanotubes in hyphenated flow injection-flame atomic absorption spectrometry system [J].
Bajaj, Sonam ;
Jain, Vanshika ;
Sharma, Niharika ;
Tiwari, Shelja ;
Saxena, Reena .
JOURNAL OF CHROMATOGRAPHY A, 2021, 1638
[43]   On-line preconcentration system for cobalt determination in bee honey using flow injection-flame AAS [J].
Cerutti, S ;
Orsi, RF ;
Kaplan, M ;
Gásquez, JA ;
Olsina, RA ;
Martinez, LD .
INSTRUMENTATION SCIENCE & TECHNOLOGY, 2004, 32 (04) :401-412
[44]   Design of a field flow preconcentration system for cadmium determination in seawater by flow-injection-atomic absorption spectrometry [J].
Yebra, MC ;
García, A ;
Carro, N ;
Moreno-Cid, A ;
Puig, L .
TALANTA, 2002, 56 (04) :777-785
[45]   On-line preconcentration and determination of chromium in parenteral solutions by flow injection - flame atomic absorption spectrometry [J].
Wuilloud, GM ;
Wuilloud, RG ;
de Wuilloud, JCA ;
Olsina, RA ;
Martinez, LD .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 31 (01) :117-124
[46]   COPRECIPITATION OF TRACE AMOUNTS OF SILICON WITH ALUMINUM HYDROXIDE AND THE DETERMINATION BY FLAME ATOMIC ABSORPTION SPECTROMETRY [J].
Shokrollahi, Ardeshir ;
Gohari, Masoud .
QUIMICA NOVA, 2014, 37 (10) :1624-1628
[48]   Flame atomic absorption spectrometry with flow-injection on-line adsorption preconcentration using a knotted reactor for cadmium determination in aqueous samples [J].
Garrido, I ;
Soto, RM ;
Carlosena, A ;
López-Mahía, P ;
Muniategui, S ;
Prada, D .
ANALYTICAL LETTERS, 2001, 34 (10) :1763-1779
[49]   On-line solid phase extraction system for chromium determination in water samples by flow injection-flame atomic absorption spectrometry [J].
Saracoglu, S ;
Soylak, M ;
Elci, L .
ANALYTICAL LETTERS, 2002, 35 (09) :1519-1530
[50]   ONLINE MICROWAVE DIGESTION OF SLURRY SAMPLES WITH DIRECT FLAME ATOMIC-ABSORPTION SPECTROMETRIC ELEMENTAL DETECTION [J].
HASWELL, SJ ;
BARCLAY, D .
ANALYST, 1992, 117 (02) :117-120